Assessing the radiofrequency shielding effect of titanium mesh on diffusion-weighted imaging: a comparative study of the twice-refocused spin-echo and Stejskal-Tanner sequences.
Imamura, Eizaburo; Jomoto, Wataru; Takatsu, Yasuo; Enoki, Takuya; Wakayama, Tsukasa; Kotoura, Noriko.
Afiliação
  • Imamura E; Department of Radiological Technology, Hyogo Medical University Hospital, 1-1 Mukogawa-cho, Nishinomiya, Hyogo, 663-8501, Japan.
  • Jomoto W; Department of Radiological Technology, Hyogo Medical University Hospital, 1-1 Mukogawa-cho, Nishinomiya, Hyogo, 663-8501, Japan. wataru-j@hyo-med.ac.jp.
  • Takatsu Y; Molecular Imaging, School of Medical Sciences, Fujita Health University, 1-98, Dengakugakubo, Kutsukake-cho, Toyoake, Aichi, 470-1192, Japan.
  • Enoki T; Department of Radiological Technology, Hyogo Medical University Hospital, 1-1 Mukogawa-cho, Nishinomiya, Hyogo, 663-8501, Japan.
  • Wakayama T; Department of Radiological Technology, Hyogo Medical University Hospital, 1-1 Mukogawa-cho, Nishinomiya, Hyogo, 663-8501, Japan.
  • Kotoura N; Department of Radiological Technology, Hyogo Medical University Hospital, 1-1 Mukogawa-cho, Nishinomiya, Hyogo, 663-8501, Japan.
Phys Eng Sci Med ; 47(3): 1051-1057, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38771443
ABSTRACT
This study compared twice-refocused spin-echo sequence (TRSE) and Stejskal-Tanner sequence (ST) to evaluate their respective effects on the image quality of magnetic resonance (MR) diffusion-weighted imaging in the presence of radiofrequency (RF) shielding effect of titanium mesh in cranioplasty. A 1.5-T MR scanner with a Head/Neck coil 20 channels and a phantom simulating the T2 and apparent diffusion coefficient (ADC) value of the human brain were used. Imaging was performed with and without titanium mesh placed on the phantom in TRSE and ST, and normalized absolute average deviation (NAAD), Dice similarity coefficient (DSC), and ADC values were calculated. The NAAD values were significantly lower for TRSE than for ST in the area below the titanium mesh, and the drop rates due to titanium mesh were 14.1% for TRSE and 9.8% for ST. The DSC values were significantly lower for TRSE than for ST. The ADC values were significantly higher for TRSE than for ST without titanium mesh. The ADC values showed no significant difference between TRSE and ST with titanium mesh. The ST had a lower RF shielding effect of titanium mesh than the TRSE.