Quantification of C1 inhibitor activity using a chromogenic automated assay: analytical and clinical performances.
Renaudineau, Yves; Sailler, Laurent; Puissant-Lubrano, Bénédicte.
Afiliação
  • Renaudineau Y; Immunology Department Laboratory, Referral Medical Biology Laboratory, 36760 Institut Fédératif de Biologie, Toulouse University Hospital Center , Toulouse, France.
  • Sailler L; 36760 Internal Medicine, University Toulouse III , Toulouse, France.
  • Puissant-Lubrano B; 36760 Competence Centre CREAK, Toulouse University Hospital Center , Toulouse, France.
Clin Chem Lab Med ; 62(12): 2477-2484, 2024 Nov 26.
Article em En | MEDLINE | ID: mdl-38861040
ABSTRACT

OBJECTIVES:

The quantification of functional C1 inhibitor activity (fC1-INH) is an important tool to diagnose bradykinin-mediated angioedema (AE), whether hereditary or acquired. For that an accurate assay is necessary, therefore we evaluated the analytical performances of a fC1-INH chromogenic assay (Berichrom®, Siemens) performed utilizing an Optilite turbidimeter (Binding Site).

METHODS:

fC1-INH was quantified by means of the chromogenic assay Berichrom®. Internal quality controls were used to determine the precision of the assay. Stability under various storage and matrix conditions, uncertainty, linearity, interference (of hemolysis, lipemia, and icterus), agreement with the manual Technochrom® assay, and diagnostic performances were further evaluated on samples from patients and healthy donors.

RESULTS:

The fC1-INH Berichrom® assay presented good performances regarding intra- and inter-assay precision (CV 1.3-4.5 % and 3.0-6.0 %, respectively), expanded uncertainty (5.5 % at normal level and 12.5 % at the clinical threshold) and linearity (rho2>0.99 range 7-130 % activity). Addition of interfering substances (hemoglobin <16 g/L, intralipid® <12 g/L, and bilirubin <1 g/L) did not affect fC1-INH quantification. fC1-INH activity from healthy donors remained stable in citrate whole blood until 4 days at room temperature, and 7 days when plasma was collected. Agreement between the automated Berichrom® assay and the manual Technochrom® assay (n=47) was excellent as obtained with both quantitative (Deming regression and Bland-Altman difference plot) and qualitative (Kappa index=1) analyses. Finally, the diagnostic performance of the quantification of fC1-INH for AE evaluated on 81 patients revealed a sensitivity of 100 %, a specificity of 97.2 %, a positive predictive value of 83.3 % and a negative predictive value of 100 %.

CONCLUSIONS:

The automated fC1-INH Berichrom® assay showed good performance, both at the analytical and diagnostic/clinical levels that allowed its usage in a clinical laboratory for C1-INH-dependent bradykinin-mediated AE research in combination with quantitative C1-INH and C4 determinations.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Proteína Inibidora do Complemento C1 Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Chem Lab Med / Clin. chem. lab. med / Clinical chemistry and laboratory medicine Assunto da revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2024 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Proteína Inibidora do Complemento C1 Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Chem Lab Med / Clin. chem. lab. med / Clinical chemistry and laboratory medicine Assunto da revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2024 Tipo de documento: Article País de afiliação: França