Biomechanical Comparison of Corticopedicular Spine Fixation versus Pedicle Screw Fixation in a Lumbar Degenerative Spondylolisthesis Finite Element Analysis Model.
Farooqi, Ali S; Narayanan, Rajkishen; Canseco, Jose A; Vaccaro, Alexander R.
Afiliação
  • Farooqi AS; Department of Orthopaedic Surgery, Rothman Orthopaedic Institute, Philadelphia, Pennsylvania, USA.
  • Narayanan R; Department of Orthopaedic Surgery, Rothman Orthopaedic Institute, Philadelphia, Pennsylvania, USA. Electronic address: Rajkishen.Narayanan@gmail.com.
  • Canseco JA; Department of Orthopaedic Surgery, Rothman Orthopaedic Institute, Philadelphia, Pennsylvania, USA.
  • Vaccaro AR; Department of Orthopaedic Surgery, Rothman Orthopaedic Institute, Philadelphia, Pennsylvania, USA.
World Neurosurg ; 2024 Jul 14.
Article em En | MEDLINE | ID: mdl-39004177
ABSTRACT

OBJECTIVE:

To compare the stability of a corticopedicular posterior fixation (CPPF) device with traditional pedicle screws for decompression and fusion in adult degenerative lumbar spondylolisthesis.

METHODS:

Finite element analysis (FEA) was used in a validated model of grade 1 L4-L5 spondylolisthesis to compare segmental stability after laminectomy alone, laminectomy with pedicle screw fixation, or laminectomy with CPPF device fixation. A 500-N follower load was applied to the model and different functional movements were simulated by applying a 7.5-Nm force in different directions. Outcomes included degrees of motion, tensile forces experienced in the CPPF device, and stresses in surrounding cortical bone.

RESULTS:

At maximum loading, laminectomy alone demonstrated a 1° increase in flexion range of motion, from 6.35° to 7.39°. Laminectomy with pedicle screw fixation and CPPF device fixation both reduced spinal segmental motion to ≤1° at maximum loading in all ranges of motion, including flexion (0.94° and 1.09°), extension (-0.85° and -1.08°), lateral bending (-0.56° and -0.96°), and torsion (0.63° and 0.91°), respectively. There was no significant difference in segmental stability between pedicle screw fixation and CPPF device fixation during maximum loading, with a difference of ≤0.4° in any range of motion. Tensile forces in the CPPF device remained ≤51% the ultimate load to failure (487 N) and stress in surrounding cortical bone remained ≤84% the ultimate stress of cortical bone (125.4 MPa) during maximum loading.

CONCLUSIONS:

CPFF fixation demonstrated similar segmental stability to traditional pedicle screw fixation whereas tensile forces and stress in surrounding cortical bone remained below the load to failure.