Your browser doesn't support javascript.
loading
RANDOMIZED TRIAL COMPARING MICROSERRATED VERSUS CONVENTIONAL INTERNAL LIMITING MEMBRANE FORCEPS FOR INTERNAL LIMITING MEMBRANE PEELING.
Starr, Matthew R; Hinkle, John C; Patel, Luv G; Ammar, Michael J; Soares, Rebecca R; Patel, Samir N; Cohen, Michael N; Hsu, Jason; Yonekawa, Yoshihiro; Ho, Allen C; Regillo, Carl D; Gupta, Omesh P.
Afiliação
  • Starr MR; Department of Ophthalmology, Mayo Clinic, Rochester, Minnesota; and.
  • Hinkle JC; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Patel LG; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Ammar MJ; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Soares RR; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Patel SN; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Cohen MN; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Hsu J; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Yonekawa Y; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Ho AC; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Regillo CD; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Gupta OP; The Retina Service, Wills Eye Hospital, Thomas Jefferson University, Philadelphia, Pennsylvania.
Retina ; 44(8): 1323-1328, 2024 Aug 01.
Article em En | MEDLINE | ID: mdl-39047125
ABSTRACT

PURPOSE:

To evaluate anatomic outcomes and surgeon response following the use of microserrated (Sharkskin, Alcon, Forth Worth, TX) internal limiting membrane (ILM) forceps compared with conventional (Grieshaber; Alcon) ILM forceps for peeling of the ILM.

METHODS:

Patients were prospectively assigned in a 11 randomized fashion to undergo ILM peeling using microserrated forceps or conventional forceps. Rates of retinal hemorrhages, deep retinal grasps, ILM regrasping, time to ILM removal, and surgeon questionnaire comparing the use of microserrated and conventional ILM forceps were analyzed.

RESULTS:

A total of 90 eyes of 90 patients were included in this study. The mean number of deep retinal grasps was higher in the conventional forceps group (1.51 ± 1.70 vs. 0.33 ± 0.56, respectively [P < 0.0001]). The mean number of failed ILM grasps was higher with conventional forceps (6.62 ± 3.51 vs. 5.18 ± 2.06 [P = 0.019]). Microserrated forceps provided more comfortability (lower number) in initiating the ILM flap (2.16 ± 0.85 vs. 1.56 ± 0.76, P < 0.001), comfortability in regrasping the ILM flap (2.51 ± 1.01 vs. 1.98 ± 0.89, P = 0.01), and comfortability in completing the ILM flap (2.42 ± 1.03 vs. 1.84 ± 1.02, P = 0.01).

CONCLUSION:

Surgeons utilizing the microserrated forceps experienced fewer deep retina grasps and fewer failed ILM grasps compared with conventional ILM forceps. The microserrated forceps was also a more favorable experience subjectively among the surgeons.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Membrana Basal / Vitrectomia / Acuidade Visual Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Retina Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Membrana Basal / Vitrectomia / Acuidade Visual Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Retina Ano de publicação: 2024 Tipo de documento: Article