McCauley, J C; Allen, J R; Selby, A M; Gaskin, K J.
Afiliação
  • McCauley JC; James Fairfax Institute of Paediatric Clinical Nutrition, Royal Alexandra Hospital for Children, Camperdown, Sydney, Australia.
Australas Phys Eng Sci Med ; 16(3): 129-36, 1993 Sep.
Article em En | MEDLINE | ID: mdl-8240141
ABSTRACT
An indirect calorimeter was developed to measure resting energy expenditure (REE) and respiratory quotient (RQ) in a clinical paediatric population. The design is suitable for any child spontaneously breathing room air. Measurements are possible from small babies to large hypermetabolic adolescents due to the ability to set and control flow from the patient headbox across the range 2-40 l/min. The calorimeter is mobile, it is electrically safe and the patient interface may be disinfected. An alcohol burning technique, that can emulate a clinical test for children of various sizes, is used to check the system, which has a high level of accuracy across a range of conditions. The indirect calorimeter has been successfully used to measure a large number of children of different sizes.
Assuntos
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Calorimetria Indireta Limite: Adolescent / Child / Child, preschool / Humans / Infant Idioma: En Revista: Australas Phys Eng Sci Med Ano de publicação: 1993 Tipo de documento: Article País de afiliação: Austrália
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Calorimetria Indireta Limite: Adolescent / Child / Child, preschool / Humans / Infant Idioma: En Revista: Australas Phys Eng Sci Med Ano de publicação: 1993 Tipo de documento: Article País de afiliação: Austrália