1.
Clin Neuroradiol ; 27(3): 371-378, 2017 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-26869445

RESUMO

PURPOSE: Purpose of our research was the evaluation of a dedicated dental surface coil in comparison with a standard head and neck coil for the improvement of dental magnetic resonance imaging (MRI). MATERIALS AND METHODS: Axial T1-weighted spin echo MRI was performed by using a newly developed dental coil for MRI and a standard head and neck coil on five volunteers. In addition, MRI was implemented with dental coil on five patients. Using the Wilcoxon test, we compared the volunteers' signal-to-noise ratio (SNR) of a variety of anatomical structures (e.g., hard tooth tissue, pulp tissue, bone, muscle tissue). Also subjective evaluation of image quality was performed on both volunteers and patients. RESULTS: Compared with the head and neck coil, the mean SNR was 3.5-fold higher on an average with the dental coil (range: from 2.7 [masseter muscle] to 4.6 [pulp tissue]). That difference was statistically significant for all evaluated structures. The higher SNR also resulted in a superior image quality as determined by subjective evaluation. CONCLUSION: Dental MRI benefits profoundly from using a dedicated dental coil.


Assuntos
Implantes Dentários , Imageamento por Ressonância Magnética/instrumentação , Adolescente , Adulto , Idoso , Polpa Dentária/diagnóstico por imagem , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Imagens de Fantasmas , Articulação Temporomandibular/diagnóstico por imagem , Adulto Jovem
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA