Magn Reson Med ; 76(4): 1314-24, 2016 10.
Artigo em Inglês | MEDLINE | ID: mdl-26480019

RESUMO

PURPOSE: The design of RF coils for MRI transmit becomes increasingly challenging at high frequencies required for MRI at 7T and above. Our goal is to show a proof of principle of a new type of transmit coil for higher field strengths. METHOD: We demonstrate an alternative transmit coil design based on dielectric waveguide principles which transfers energy via evanescent wave coupling. The operating principles and conditions are explored by simulations. The waveguide is applied for in vivo imaging at 7T. RESULTS: The waveguide can be an efficient transmit coil when four conditions are fulfilled: (1) the waveguide should be operated just above the cutoff frequency of the lowest order transverse electric mode, (2) the waveguide should not operate at a frequency where the wavelength fits an integer number of times in the waveguide length and standing wave patterns become very prominent, (3) for homogeneous excitation, the waveguide should be bent around the object, and (4) there should be an air gap between the waveguide and the object. CONCLUSIONS: By choosing the dielectric and the dimensions adequately, the dielectric waveguide couples the magnetic field efficiently into the body. The waveguide can be redesigned for higher frequencies by simple adaptations and may be a promising transmit alternative. Magn Reson Med 76:1314-1324, 2016. © 2015 Wiley Periodicals, Inc.


Assuntos
Aumento da Imagem/instrumentação , Imageamento por Ressonância Magnética/instrumentação , Magnetismo/instrumentação , Transdutores , Desenho Assistido por Computador , Impedância Elétrica , Transferência de Energia , Desenho de Equipamento , Análise de Falha de Equipamento , Imagens de Fantasmas , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
2.
Magn Reson Med ; 73(5): 2038-46, 2015 May.
Artigo em Inglês | MEDLINE | ID: mdl-24903409

RESUMO

PURPOSE: Neoadjuvant treatment response in lymph nodes predicts patient outcome, but existing methods do not track response during therapy accurately. In this study, specialized hardware was used to adapt high-field (7T) (31) P magnetic resonance spectroscopy (MRS), which has been shown to track treatment response in small breast tumors, to monitor axillary lymph nodes. METHOD: A dual-tuned quadrature coil that is a (31) P (120 MHz) transceiver and a (1) H (300 MHz) receiver was designed using a novel detune circuit. The transceiver/receiver coil in the axilla is used with a fractionated dipole antenna on the back of the subject and the conventional breast coil for transmit. RESULTS: The novel circuit detuned the (1) H resonance without disturbing the (31) P resonance. In vivo demonstrations included: >80% homogeneous B1 (+) for (1) H over the axilla, identification of a small (3-mm diameter) lymph node, and (31) P MR spectra from a single healthy lymph node. The setup can detect <2 millimolar concentrations of metabolites from a 2-mL voxel. CONCLUSIONS: The first (31) P MR spectrum from an in vivo lymph node indicates that the presented design may be sufficiently sensitive to detect metabolic response to neoadjuvant therapy. Multinuclei MRS of the lymph nodes at 7T is possible through combining lightweight antenna elements with dual-tuned transceiver/receive-only coils.


Assuntos
Neoplasias da Mama/diagnóstico , Neoplasias da Mama/terapia , Metástase Linfática/diagnóstico , Imageamento por Ressonância Magnética/instrumentação , Espectroscopia de Prótons por Ressonância Magnética/instrumentação , Adulto , Axila , Desenho de Equipamento , Segurança de Equipamentos , Feminino , Humanos , Linfonodos/patologia , Metástase Linfática/patologia , Terapia Neoadjuvante , Imagens de Fantasmas , Prognóstico , Valores de Referência , Sensibilidade e Especificidade
3.
MAGMA ; 28(1): 49-56, 2015 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-24862451

RESUMO

OBJECTIVE: To identify a shielding material compatible with optical head-motion tracking for prospective motion correction and which minimizes radio frequency (RF) radiation losses at 7 T without sacrificing line-of-sight to an imaging target. MATERIALS AND METHODS: We evaluated a polyamide mesh coated with silver. The thickness of the coating was approximated from the composition ratio provided by the material vendor and validated by an estimate derived from electrical conductivity and light transmission measurements. The performance of the shield is compared to a split-copper shield in the context of a four-channel transmit-only loop array. RESULTS: The mesh contains less than a skin-depth of silver coating (300 MHz) and attenuates light by 15 %. Elements of the array vary less in the presence of the mesh shield as compared to the split-copper shield indicating that the array behaves more symmetrically with the mesh shield. No degradation of transmit efficiency was observed for the mesh as compared to the split-copper shield. CONCLUSION: We present a shield compatible with future integration of camera-based motion-tracking systems. Based on transmit performance and eddy-current evaluations the mesh shield is appropriate for use at 7 T.


Assuntos
Imageamento por Ressonância Magnética/instrumentação , Dispositivos Ópticos , Proteção Radiológica/instrumentação , Ondas de Rádio , Transdutores , Desenho de Equipamento , Análise de Falha de Equipamento
4.
PLoS One ; 14(3): e0213107, 2019.
Artigo em Inglês | MEDLINE | ID: mdl-30830934

RESUMO

BACKGROUND: The quality of carotid wall MRI can benefit substantially from a dedicated RF coil that is tailored towards the human neck geometry and optimized for image signal-to-noise ratio (SNR), parallel imaging performance and RF penetration depth and coverage. In last decades, several of such dedicated carotid coils were introduced. However, a comparison of the more successful designs is still lacking. OBJECTIVE: To perform a head-to-head comparison over four dedicated MR carotid surface coils with 4, 6, 8 and 30 coil elements, respectively. MATERIAL AND METHODS: Ten volunteers were scanned on a 3T scanner. For each subject, multiple black-blood carotid vessel wall images were measured using the four coils with different parallel imaging settings. The performance of the coils was evaluated and compared in terms of image coverage, penetration depth and noise correlations between elements. Vessel wall of a common carotid section was delineated manually. Subsequently, images were assessed based on vessel wall morphology and image quality parameters. The morphological parameters consisted of the vessel wall area, thickness, and normalized wall index (wall area/total vessel area). Image quality parameters consisted of vessel wall SNR, wall-lumen contrast-to-noise ratio (CNR), the vessel g-factor, and CNRindex ((wall-lumen signal) / (wall+lumen signal)). Repeated measures analysis of variance (rmANOVA) was applied for each parameter for the averaged 10 slices for all volunteers to assess effect of coil and SENSE factor. If the rmANOVA was significant, post-hoc comparisons were conducted. RESULTS: No significant coil effect were found for vessel wall morphological parameters. SENSE acceleration affected some morphological parameters for 6- and 8-channel coils, but had no effect on the 30-channel coil. The 30-channel coil achieved high acceleration factors (10x) with significantly lower vessel g-factor values (ps ≤ 0.01), but lower vessel wall SNR and CNR values (ps ≤ 0.01). CONCLUSION: All four coils were capable of high-quality carotid MRI. The 30-channel coil is recommended when rapid image acquisition acceleration is required for 3D measurements, whereas 6- and 8-channel coils demonstrated the highest SNR performance.


Assuntos
Artérias Carótidas/anatomia & histologia , Imageamento por Ressonância Magnética/instrumentação , Intensificação de Imagem Radiográfica/métodos , Adulto , Algoritmos , Artérias Carótidas/diagnóstico por imagem , Feminino , Voluntários Saudáveis , Humanos , Masculino , Imagens de Fantasmas , Razão Sinal-Ruído , Adulto Jovem
SELEÇÃO DE REFERÊNCIAS