RESUMO
PURPOSE: In radiation therapy, surface guidance can be used for patient setup and intra-fraction motion monitoring. The surface guided radiation therapy (SGRT) system from Varian Medical systems, IDENTIFYTM, consists of three pods, including cameras and a random pattern projector, mounted on the ceiling. The information captured by the cameras is used to make a reconstruction of the surface. The aim of the study was to assess the technical performance of this SGRT system on a closed-bore linac. METHODS: Phantom measurements were performed to assess the accuracy, precision, reproducibility and temporal stability of the system, both in align and in load position. Translations of the phantoms in lateral, longitudinal, and vertical direction, and rotations around three axes (pitch, roll and yaw) were performed with an accurate, in-house built, positioning stage. Different phantom geometries and different surface colors were used, and various ambient light intensities were tested. RESULTS: The accuracy of the IDENTIFYTM system at the closed-bore linac was 0.07 mm and 0.07 degrees at load position, and 0.06 mm and 0.01 degrees at align position for the white head phantom. The precision was 0.02 mm and 0.02 degrees in load position, and 0.01 mm and 0.02 degrees in align position. The accuracy for the Penta-Guide phantom was comparable to the white head phantom, with 0.06 mm and 0.01 degrees in align position. The system was slightly less accurate for translations of the CIRS lung phantom in align position (0.20 mm, 0.05 degrees). Reproducibility measurements showed a variation of 0.02 mm in load position. Regarding the temporal stability, the maximum drift over 30 min was 0.33 mm and 0.02 degrees in load position. No effect of ambient light level on the accuracy of the IDENTIFYTM system was observed. Regarding different surface colors, the accuracy of the system for a black phantom was slightly worse compared to a white surface, but not clinical relevant. CONCLUSION: The IDENTIFYTM system can adequately be used for motion monitoring on a closed-bore linac with submillimeter accuracy. The results of the performed measurements meet the clinical requirements described in the guidelines of the AAPM and the ESTRO.
Assuntos
Radioterapia Guiada por Imagem , Humanos , Reprodutibilidade dos Testes , Radioterapia Guiada por Imagem/métodos , Imagens de Fantasmas , Posicionamento do Paciente/métodos , Planejamento da Radioterapia Assistida por Computador/métodosRESUMO
OBJECTIVE: Surface-guided radiation therapy (SGRT, AlignRT) was used to analyze motion during stereotactic body radiotherapy (SBRT) in lung cancer patients and to explore the margin of the planning target volume (PTV). METHODS: The residual errors of the AlignRT were evaluated based on grayscale cone-beam computed tomography registration results before each treatment. AlignRT log file was used to analyze the correlation between the frequency and longest duration of errors larger than 2 mm and lasting longer than 2 s and maximum error with age and treatment duration. The displacement value at the end of treatment, the average displacement value, and the 95% probability density displacement interval were defined as intrafraction errors, and PTV1, PTV2, PTV3 were calculated by Van Herk formula or Z score analysis. Organ dosimetric differences were compared after the experience-based margin was replaced with PTV3. RESULTS: The interfraction residual errors were Vrt0 , 0.06 ± 0.18 cm; Lng0 , -0.03 ± 0.19 cm; Lat0 , 0.02 ± 0.15 cm; Pitch0 , 0.23 ± 0.7°; Roll0 , 0.1 ± 0.69°; Rtn0 , -0.02 ± 0.79°. The frequency, longest duration and maximum error in vertical direction were correlated with treatment duration (r = 0.404, 0.353, 0.283, p < 0.05, respectively). In the longitudinal direction, the frequency was correlated with age and treatment duration (r = 0.376, 0.283, p < 0.05, respectively), maximum error was correlated with age (r = 0.4, P < 0.05). Vertical, longitudinal, lateral margins of PTV1, PTV2, PTV3 were 2 mm, 4 mm, 2 mm; 2 mm, 2 mm, 2 mm, 3 mm, 5 mm, 3 mm, respectively. After replacing the original PTV, mean lung dose (MLD), 2-cm3 chest wall dose (CD), lung V20 decreased by 0.2 Gy, 2.1 Gy, 0.5%, respectively (p < 0.05). CONCLUSION: AlignRT can be used for interfraction setup and monitoring intrafraction motion. It is more reasonable to use upper and lower limits of the 95% probability density interval as an intrafraction error.
Assuntos
Neoplasias Pulmonares , Radiocirurgia , Radioterapia Guiada por Imagem , Humanos , Neoplasias Pulmonares/radioterapia , Neoplasias Pulmonares/cirurgia , Radiocirurgia/métodos , Radioterapia Guiada por Imagem/métodos , Pulmão , Tomografia Computadorizada de Feixe Cônico , Planejamento da Radioterapia Assistida por Computador/métodos , Erros de Configuração em Radioterapia/prevenção & controleRESUMO
PURPOSE: The use of volumetric modulated arc therapy (VMAT), simultaneous integrated boost (SIB), and hypofractionated regimen requires adequate patient setup accuracy to achieve an optimal outcome. The purpose of this study was to assess the setup accuracy of patients receiving left-sided breast cancer radiotherapy using deep inspiration breath-hold technique (DIBH) and surface guided radiotherapy (SGRT) and to calculate the corresponding setup margins. METHODS: The patient setup accuracy between and within radiotherapy fractions was measured by comparing the 6DOF shifts made by the SGRT system AlignRT with the shifts made by kV-CBCT. Three hundred and three radiotherapy fractions of 23 left-sided breast cancer patients using DIBH and SGRT were used for the analysis. All patients received pre-treatment DIBH training and visual feedback during DIBH. An analysis of variance (ANOVA) was used to test patient setup differences for statistical significance. The corresponding setup margins were calculated using the van Herk's formula. RESULTS: The intrafractional patient setup accuracy was significantly better than the interfractional setup accuracy (p < 0.001). The setup margin for the combined inter- and intrafractional setup error was 4, 6, and 4 mm in the lateral, longitudinal, and vertical directions if based on SGRT alone. The intrafractional error contributed ≤1 mm to the calculated setup margins. CONCLUSION: With SGRT, excellent intrafractional and acceptable interfractional patient setup accuracy can be achieved for the radiotherapy of left-sided breast cancer using DIBH and modern radiation techniques. This allows for reducing the frequency of kV-CBCTs, thereby saving treatment time and radiation exposure.
Assuntos
Suspensão da Respiração , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Erros de Configuração em Radioterapia , Radioterapia Guiada por Imagem , Radioterapia de Intensidade Modulada , Neoplasias Unilaterais da Mama , Humanos , Feminino , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos , Neoplasias Unilaterais da Mama/radioterapia , Erros de Configuração em Radioterapia/prevenção & controle , Radioterapia Guiada por Imagem/métodos , Órgãos em Risco/efeitos da radiação , Pessoa de Meia-Idade , Neoplasias da Mama/radioterapia , PrognósticoRESUMO
PURPOSE: A markerless workflow for the treatment of breast cancer patients has been introduced and evaluated retrospectively. It includes surface-guided radiation therapy (SGRT)-only positioning for patients with small cone beam CT (CBCT) position corrections during the first five fractions. Prerequisites and the frequency of its clinical application were evaluated, as well as potential benefits in terms of treatment time and dose savings, the frequency of CBCT scans, and the accuracy of the positioning. METHODS: A group of 100 patients treated with the new workflow on two Versa HD linacs has been compared to a matched control group of patients treated with the former workflow, which included prepositioning with skin markings and lasers, SGRT and daily CBCT. The comparison was based on the evaluation of logfiles. RESULTS: Of the patients treated with the new workflow, 40% did not receive daily CBCT scans. This resulted in mean time savings of 97â¯s, 166â¯s and 239â¯s per fraction for the new workflow, for patients treated without daily CBCT and for SGRT-only fractions, respectively, when compared to the old workflow. Dose savings amounted to a weighted computed tomography dose index reduction of CTDIWâ¯= 2.56 cGy on average for normofractionated treatment and weekly CBCTs, while for patients not treated with daily CBCT, SGRT-based positioning accuracy was 5.2â¯mm for the mean translational magnitude, as evaluated by CBCT. CONCLUSION: For 40% of the patients, after five fractions with small CBCT corrections, the workflow could be changed to SGRT-only positioning with weekly CBCT. This leads to imaging dose and time savings and thus also reduced intrafraction motion, potentially increased patient throughput and patient comfort, while assuring appropriate positioning accuracy.
Assuntos
Neoplasias da Mama , Radioterapia Guiada por Imagem , Humanos , Feminino , Neoplasias da Mama/diagnóstico por imagem , Neoplasias da Mama/radioterapia , Posicionamento do Paciente/métodos , Fluxo de Trabalho , Estudos Retrospectivos , Radioterapia Guiada por Imagem/métodos , Tomografia Computadorizada de Feixe Cônico/métodos , Planejamento da Radioterapia Assistida por Computador/métodosRESUMO
PURPOSE: We evaluate an SGRT device (Voxelan HEV-600 M/RMS) installed with Radixact, with the view angle of the Voxelan's camera at 74 degrees. The accuracy of Voxelan with this steep angle was evaluated with phantom experiments and inter-fractional setup errors of patients. METHODS: In the phantom experiments, the difference between the measured values of Voxelan from the truth was evaluated for translations and rotations. The inter-fractional setup error between the setup using skin markers with laser localizer (laser setup: LS) and the setup using Voxelan (surface setup: SS) was compared for head and neck (N = 19), chest (N = 7) and pelvis (N = 9) cases. The inter-fractional setup error was calculated by subtracting from bone matching by megavoltage computed tomography (MVCT) as ground truth. RESULTS: From the phantom experiments, the average difference between the measured values of Voxelan from the truth was within 1 mm and 1 degree. In all cases, inter-fractional setup error based on MVCT was not significantly different between LS and SS by Welch's t-test (P > 0.05). The vector offset of the LS for head and neck, chest, and pelvis were 6.5, 9.6, and 9.6 mm, respectively, and that of the SS were 5.8, 8.6, and 12.6 mm, respectively. Slight improvement was observed for the head and neck, and chest cases, however, pelvis cases were not improved because the umbilical region could not be clearly visualized as a reference. CONCLUSION: The results show that SS in Voxelan with an installation angle of 74 degrees is equal to or better than LS.
Assuntos
Planejamento da Radioterapia Assistida por Computador , Radioterapia Guiada por Imagem , Humanos , Planejamento da Radioterapia Assistida por Computador/métodos , Cabeça/diagnóstico por imagem , Pescoço , Tomografia Computadorizada de Feixe Cônico , Tórax/diagnóstico por imagem , Radioterapia Guiada por Imagem/métodosRESUMO
PURPOSE: To commission and assess the clinical performance of a new commercial surface imaging (SI) system by analyzing intra-fraction motion from the initial cohort of patients treated with frameless stereotactic radiosurgery (fSRS). METHODS: The IDENTIFYTM SI system was commissioned for clinical use on an Edge (Varian Medical Systems, Palo Alto, CA) linear accelerator. All patients who received intracranial radiotherapy with HyperArcTM (Varian Medical Systems, Palo Alto, CA) were immobilized with the EncompassTM (Qfix, Avondale, PA) thermoplastic mask and monitored for intra-fraction motion with SI. IDENTIFYTM log files were correlated with trajectory log files to correlate treatment parameters with SI-reported offsets. IDENTIFYTM reported offsets were correlated with gantry and couch angles to assess system performance for obstructed and clear camera field of view. Data were stratified by race to evaluate performance differences due to skin tone. RESULTS: All commissioning data were found to meet recommended tolerances. IDENTIFYTM was used to monitor intra-fraction motion on 1164 fractions from 386 patients. The median magnitude of translational SI reported offsets at the end of treatment was 0.27 mm. SI reported offsets were shown to increase when camera pods are blocked by the gantry with larger increases seen at non-zero couch angles. With camera obstruction, the median magnitude of the SI reported offset was 0.50 and 0.80 mm for White and Black patients, respectively. CONCLUSIONS: IDENTIFYTM performance during fSRS is comparable to other commercially available SI systems where offsets are shown to increase at non-zero couch angles and during camera pod blockage.
Assuntos
Neoplasias Encefálicas , Radiocirurgia , Radioterapia de Intensidade Modulada , Humanos , Radiocirurgia/métodos , Posicionamento do Paciente/métodos , Aceleradores de Partículas , Imagens de Fantasmas , Radioterapia de Intensidade Modulada/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Neoplasias Encefálicas/diagnóstico por imagem , Neoplasias Encefálicas/cirurgia , Neoplasias Encefálicas/radioterapiaRESUMO
PURPOSE/OBJECTIVE(S): Surface-guided radiation therapy (SGRT) can track the patient surface noninvasively to complement radiographic image-guided radiation therapy with a standard 3-camera system and a single radiation/image isocenter. Here we report the commissioning of a novel SGRT system that monitors three imaging isocenters locations in a proton half-gantry room with a unique 5-camera configuration. MATERIALS/METHODS: The proton half-gantry room has three image isocenters, designated ISO-0, ISO-1, and ISO-2, to cover various anatomical sites via a robotic ceiling-mounted cone-beam CT. Although ISO-0 and ISO-1 are used to image the cranium, head and neck, and thoracic regions, ISO-2 is often used to image body and extremity sites and contiguous craniospinal target volumes. The five-camera system was calibrated to the radiographic isocenter by using a stereotactic radiosurgery cube phantom for each image isocenter. RESULTS: The performance of this 5-camera system was evaluated for 6 degrees of freedom in three categories: (1) absolute setup accuracy relative to the radiographic kV image isocenter based on the DICOM reference; (2) relative shift accuracy based on a reference surface capture; and (3) isocenter tracking accuracy from one isocenter to another based on a reference surface capture. The evaluation revealed maximum deviations of 0.8, 0.2, and 0.6 mm in translation and 0.2°, 0.1°, and 0.1° in rotation for the first, second, and third categories, respectively. Comparing the dosimetry and latency with static and gated irradiation revealed a 0.1% dose difference and positional differences of 0.8 mm in X and 0.9 mm in Y with less than 50 ms temporal accuracy. CONCLUSION: The unique 5-camera system configuration provides SGRT at the treatment isocenter (ISO-0) and also imaging isocenter locations (ISO-0, ISO-1, and ISO-2) to ensure correct patient positioning before and after radiographic imaging, especially during transitions from the offset imaging isocenters to the treatment isocenter.
Assuntos
Prótons , Radioterapia Guiada por Imagem , Humanos , Imagens de Fantasmas , Radioterapia Guiada por Imagem/métodos , Tomografia Computadorizada de Feixe Cônico/métodos , Posicionamento do PacienteRESUMO
PURPOSE: The goal of this study was to investigate how the choice of the region of interest (ROI) affects the registration results of surface imaging for daily positioning of breast cancer patients. METHODS: The AlignRT system (VisionRT, London) and the XVI Cone beam CT (CBCT; Elekta, Stockholm) installed on two Versa HD linacs (Elekta) were used in this study, which included 28 patients (160 fractions). In the clinical workflow, patients were prepositioned with AlignRT and then shifted in 6 degrees of freedom (DOF) according to the CBCT. A new reference capture was taken immediately afterward. Retrospectively, the surface capture resulting from prepositioning was registered to the latest reference capture. By varying the ROI used for registration, the surface-based results were optimized in terms of minimizing the deviation to the clinically applied CBCT shifts. Two sets of ROIs were used: one obtained by applying a variable margin to the breast surface, another by combining ROIs of anatomical structures, including the sternum and contralateral breast. RESULTS: Registration results showed significant differences from one ROI to another. Generally, the results improved with increasing ROI size, especially for rotational DOFs. ROIs, including the axilla or supraclavicular lymph drainage region, did not yield an improved registration result. On the other hand, an ROI comprising the breast surface, sternum, and a belt caudal to the breasts decreased the average magnitude of the translational and rotational deviations by 6.6% and 30.8% (p < 0.01), respectively, compared to the breast surface only results. CONCLUSION: The influence of the ROI choice on surface imaging registration results was analyzed and the surface-based shifts were compared to clinically applied CBCT shifts. An optimal ROI for the treatment of breast cancer patients, consisting of the breast surface, sternum, and a belt, was identified.
Assuntos
Neoplasias da Mama , Neoplasias da Mama/diagnóstico por imagem , Neoplasias da Mama/radioterapia , Tomografia Computadorizada de Feixe Cônico , Feminino , Humanos , Planejamento da Radioterapia Assistida por Computador , Estudos RetrospectivosRESUMO
The Catalyst HD (C-RAD Positioning AB, Uppsala, Sweden) optical surface imaging (OSI) system is able to manage interfractional patient positioning, intrafractional motion monitoring, and non-contact respiratory gating without x-ray exposure for radiation therapy. In recent years, a novel high-precision surface registration algorithm for stereotactic radiosurgery (SRS algorithm) has been released. This study aimed to evaluate the technical performance of the OSI system using rigid phantoms, by comparing the conventional and SRS algorithms. To determine the system's technical performance, isocenter displacements were calculated by surface image registration via the OSI system using head, thorax, and pelvis rigid phantoms. The reproducibility of positioning was evaluated by the mean value calculated by repeating the registration 10 times, without moving each phantom. The accuracy of positioning was evaluated by the mean value of the residual error, where the 10 offset values given to each phantom were subtracted from the isocenter displacement values. The stability of motion monitoring was evaluated by measuring isocenter drift during 20 min and averaging it over 10 measurements. For the head phantom, all tests were compared with the mask types and algorithms. As a result, for all sites and both algorithms, the reproducibility, accuracy, and stability for translation and rotation were <0.1 mm and <0.1°, <1.0 mm and <1.0°, and <0.1 mm and <0.1°, respectively. In particular, the SRS algorithm had a small absolute error and standard deviation of calculated isocenter displacement, and a significantly higher reproducibility and accuracy than the conventional algorithm (P < 0.01). There was no difference in the stability between the algorithms (P = 0.0280). The SRS algorithm was found to be suitable for the treatment of rigid body sites with less deformation and small area, such as the head and face.
Assuntos
Radiocirurgia , Algoritmos , Humanos , Imagens de Fantasmas , Reprodutibilidade dos Testes , SuéciaRESUMO
Surface guided radiation therapy (SGRT) is increasingly being adopted for use in radiation treatment delivery for Head and Neck (H&N) cancer patients. This study investigated the improvement of patient setup accuracy and reduction of setup time for SGRT compared to a conventional setup. A total of 60 H&N cancer patients were retrospectively included. Patients were categorized into three groups: oral cavity, oropharynx and nasopharynx/sinonasal sites with 20 patients in each group. They were further separated into two (2) subgroups, depending on whether they were set up with the aid of SGRT. The Align-RT™ system was used for SGRT in this work. Positioning was confirmed by daily kV-kV imaging in conjunction with weekly CBCT scans. Translational and rotational couch shifts along with patient setup times were recorded. Imaging setup time, which was defined as the elapsed time from the acquisition of the first image set to the end of the last image set, was recorded. Average translational shifts were larger in the non-SGRT group. Vertical shifts showed the most significant reduction in the SGRT group for both oropharynx and oral cavity groups. Pitch corrections were significantly higher in the SGRT group for oropharynx patients and higher pitch corrections were also observed in the SGRT groups of oral cavity and nasopharynx/sinonasal patients. The average setup time when SGRT guidance was employed was shorter for all three treatment sites although this did not reach statistical significance. The largest time reduction between the SGRT and non-SGRT groups was seen in the nasopharynx/sinonasal group. This study suggests that the use of SGRT decreases the magnitude of translational couch shifts during patient setup. However, the rotational corrections needed were generally higher with SGRT group. When SGRT was employed, a definite reduction in patient setup time was observed for nasopharynx/sinonasal and hypopharynx cancer patients.
Assuntos
Neoplasias de Cabeça e Pescoço , Planejamento da Radioterapia Assistida por Computador , Radioterapia Guiada por Imagem , Neoplasias de Cabeça e Pescoço/diagnóstico por imagem , Neoplasias de Cabeça e Pescoço/radioterapia , Humanos , Estudos RetrospectivosRESUMO
Image-guided radiotherapy (IGRT) systems using ionizing radiation may increase the risk of secondary cancer and normal tissue toxicity due to additional radiation exposure caused by large field sizes or repeated scans during X-ray imaging. As an alternative to these modalities, surface-guided radiotherapy (SGRT) systems which do not employ ionizing radiation have been developed. This study presents a comprehensive performance evaluation of the Varian Identify SGRT system by using an anthropomorphic Alderson Rando phantom in three different aspects: (a) the accuracy and reproducibility of the system in different regions of interest (ROI) for varying couch displacements, (b) the setup accuracy of the system for patient positioning based on different computed tomography (CT) slice thicknesses, and (c) the potential influence of obstructing SGRT cameras by the gantry on the system's overall accuracy and reproducibility. The accuracy and reproducibility of the SGRT system fell within 1 mm and 1°. Nevertheless, in certain situations, these values were observed to exceed prescribed limits. Consequently, concerning SGRT tolerance limits for treatment applications, careful consideration of ROIs and offset values of the system is crucial. We also recommend that patients should ideally be set up during 0° gantry rotation, and the on-board imaging (OBI) system should be retracted to prevent obstruction of the cameras. Additionally, reference CT images with a slice thickness of under 3 mm are recommended for this purpose.
Assuntos
Imagens de Fantasmas , Radioterapia Guiada por Imagem , Humanos , Radioterapia Guiada por Imagem/métodos , Reprodutibilidade dos Testes , Tomografia Computadorizada por Raios XRESUMO
INTRODUCTION: Surface-guided radiation therapy (SGRT) has emerged as a powerful tool to improve patient setup accuracy in radiation therapy (RT). Combined with the goal of increasing RT accuracy is an ongoing effort to decrease RT side effects. The application of a prophylactic skin dressing to the treatment site is a well-documented method of reducing skin-related side effects from RT. This paper aims to investigate whether the application of Mepitel, a prophylactic skin dressing, has an impact on the accuracy of surface-guided patient setups in chest wall RT. METHODS: A retrospective analysis of daily image-guided Online Corrections (OLCs) from patients undergoing chest wall irradiation with SGRT was performed. Translational (superior-inferior, lateral, and anterior-posterior) OLC magnitude and direction were compared between patients treated with Mepitel applied and those treated without. Systematic and random errors were calculated and compared between groups. RESULTS: OLCs from 275 fractions were analysed. Mean OLCs were larger for patients with Mepitel applied in the superior_inferior axis (0.34 vs. 0.22 cm, P = 0.049) and for the combined translational vector (0.54 vs. 0.43 cm, P = 0.043). Combined translational systematic error was slightly larger for patients with Mepitel applied (0.15 vs. 0.09 cm). CONCLUSION: Mepitel can impact the accuracy of SGRT patient-positioning in chest wall RT. The variation however is small and unlikely to have any clinical impact if SGRT is coupled with image guidance and appropriate PTV margins. Further investigation is required to assess the effect of Mepitel on SGRT accuracy in other treatment sites, as well as any potential dosimetric impacts.
Assuntos
Bandagens , Posicionamento do Paciente , Parede Torácica , Humanos , Parede Torácica/efeitos da radiação , Masculino , Feminino , Pessoa de Meia-Idade , Idoso , Estudos Retrospectivos , Radioterapia Guiada por Imagem/métodos , Pele/efeitos da radiaçãoRESUMO
PURPOSE: The goal of this study was to obtain maximum allowed shift deviations from planning position in six degrees of freedom (DOF), that can serve as threshold values in surface guided radiation therapy (SGRT) of breast cancer patients. METHODS: The robustness of conformal treatment plans of 50 breast cancer patients against 6DOF shifts was investigated. For that, new dose distributions were calculated on shifted computed tomography scans and evaluated with respect to target volume and spinal cord dose. Maximum allowed shift values were identified by imposing dose constraints on the target volume dose coverage for 1DOF, and consecutively, for 6DOF shifts using an iterative approach and random sampling. RESULTS: Substantial decreases in target dose coverage and increases of spinal cord dose were observed. Treatment plans showed highly differing robustness for different DOFs or treated area. The sensitivity was particularly high if clavicular lymph nodes were irradiated, for shifts in lateral, vertical, roll or yaw direction, and showed partly pronounced asymmetries. Threshold values showed similar properties with an absolute value range of 0.8 mm to 5 mm and 1.4° to 5°. CONCLUSION: The robustness analysis emphasized the necessity of taking differences between DOFs and asymmetrical sensitivities into account when evaluating the dosimetric impact of position deviations. It also highlighted the importance of rotational shifts, especially if clavicular lymph nodes were irradiated. A practical approach of determining 6DOF shift limits was introduced and a set of threshold values applicable for SGRT based patient motion control was identified.
Assuntos
Neoplasias da Mama , Radioterapia Conformacional , Radioterapia Guiada por Imagem , Humanos , Feminino , Neoplasias da Mama/radioterapia , Mama , ClavículaRESUMO
This study aimed to evaluate the detection accuracy of the AlignRT-InBore system in surface-guided radiation therapy using a phantom and to determine the feasibility of the system by conducting a comparative analysis with cone-beam computed tomography (CBCT) registration. The AlignRT-InBore system integrated with the ETHOS Therapy was used. A phantom and a QUASAR phantom were employed to examine the specific areas of interest relevant to clinical cases. The evaluation involved monitoring translations for approximately 30 min and assessing the position detection accuracy for static and moving objects. Fifty clinical cases were used to evaluate the position detection accuracy and its relationship with the localization accuracy of CBCT before treatment. The detection accuracy of static and moving objects was within 1.0 mm using the phantom. However, the longitudinal direction tended to be larger than the other directions. Regarding the accuracy of localization in clinical cases, a strong and statistically significant (p < 0.01) correlation was observed in each direction. A detection accuracy within 1.0 mm is possible for static and moving objects. The detection accuracy of the patient setup using the InBore optical patient positioning system was extremely high, and the patient could be detected with high precision, suggesting its usefulness.
Assuntos
Posicionamento do Paciente , Radioterapia Guiada por Imagem , Humanos , Posicionamento do Paciente/métodos , Tomografia Computadorizada de Feixe Cônico/métodos , Radioterapia Guiada por Imagem/métodos , Imagens de Fantasmas , Algoritmos , Planejamento da Radioterapia Assistida por Computador/métodosRESUMO
A calibration phantom made of Derlin requires manual translational and rotational adjustments when calibrating a light-section-based optical surface monitoring system (VOXELAN) with a phantom material that insufficiently reflects the red-slit laser of the system. This study aimed to develop a new calibration phantom using different materials and to propose a procedure that minimizes setup errors. The new phantom, primarily made of PET100, which exhibits good reflectivity without scattering or attenuating the red-slit laser at the phantom surface, was shaped in a manner similar to that of previous designs. The detection accuracy and stability were evaluated using six different regions of interest (ROIs) and compared with previous phantom designs. The coordinate coincidence between the machine and VOXELAN was compared for both phantom designs. The detection accuracy and stability of the new phantom in the reference ROI setting were found to be better than those of previous phantoms. In the lateral, longitudinal, and vertical directions, the coordinate coincidences in translational directions for the previous phantom were obtained at 1.07 ± 0.66, 1.46 ± 0.47, and 0.26 ± 0.83 mm, whereas those for the new phantom were obtained at 0.28 ± 0.21, 0.18 ± 0.30, and - 0.30 ± 0.29 mm, respectively. The rotational errors of the two phantoms were identical. The new phantom exhibited improved detection stability because of its good reflectivity. Additionally, the new placement procedure was linked to the six-degrees-of-freedom couch. A combination of the new phantom and its new placement procedure is suitable for coordinate calibration of VOXELAN.
Assuntos
Calibragem , Imagens de FantasmasRESUMO
Surface guidance systems enable patient positioning and motion monitoring without using ionising radiation. Surface Guided Radiation Therapy (SGRT) has therefore been widely adopted in radiation therapy in recent years, but guidelines on workflows and specific quality assurance (QA) are lacking. This ESTRO-ACROP guideline aims to give recommendations concerning SGRT roles and responsibilities and highlights common challenges and potential errors. Comprehensive guidelines for procurement, acceptance, commissioning, and QA of SGRT systems installed on computed tomography (CT) simulators, C-arm linacs, closed-bore linacs, and particle therapy treatment systems are presented that will help move to a consensus among SGRT users and facilitate a safe and efficient implementation and clinical application of SGRT.
Assuntos
Radioterapia Guiada por Imagem , Humanos , Aceleradores de Partículas , Posicionamento do Paciente , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Guiada por Imagem/métodos , Tomografia Computadorizada por Raios X/métodosRESUMO
PURPOSE: Surface-guided radiotherapy is useful for the pre-positioning and monitoring of radiotherapy. The purpose of this study was to investigate the impact of surface guidance on the repeatability of patient localization and to estimate the specific point at which high positional errors occur. MATERIALS AND METHODS: Ten patients without the VOXELAN system (non-VXLN group) and 10 patients with the VOXELAN as the pre-positioning procedure (VXLN group) were included in this analysis. Twelve regions of interest (ROI) were defined in all the patients to verify any misalignment during radiotherapy. Thirteen ROIs were defined on the isocenter. RESULTS: Compared with the non-VXLN group, the translational positional errors of the VXLN group were the same for all the ROIs. The mean translational positional errors of the VXLN group in the longitudinal direction were approximately 0.1mm, and the standard deviation was the largest among the three directions in all the ROIs. The magnitude of the standard deviation in the non-VXLN group varied independently of the ROI and direction. The standard deviations of the VXLN group in the longitudinal direction were large in all the ROIs, while the standard deviations in the vertical and lateral directions were small. CONCLUSION: Pre-positioning with a surface guidance system reduced the body twist and rotation, which could not be corrected by image-guided radiotherapy alone. Since the VOXELAN can detect positioning errors quickly and without additional radiation exposure to the patient, it can be used as a tool for pre-positioning in radiotherapy.
Assuntos
Posicionamento do Paciente , Radioterapia Guiada por Imagem , Humanos , Posicionamento do Paciente/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Guiada por Imagem/métodos , RotaçãoRESUMO
Thermal drift of optical systems employed for surface guided radiation therapy (SGRT) adds uncertainty to patient setup and monitoring. This work describes methods to measure the drift of individual camera pods as well as the drift of the combined clinical signal. It presents results for four clinical C-Rad Catalyst+ HD systems. Based on the measured clinical drift, recipes are provided on how to calculate relevant uncertainties in patient setup and patient position monitoring with SGRT. Strategies to reduce the impact of drift are explained. While the results are specific to the systems investigated, the methodology is transferable and the clinical recipes are universally applicable.
RESUMO
AIM: To test inter-fraction reproducibility, intrafraction stability, technician aspects, and patient/physician's comfort of a dedicated immobilization solution for Brain Linac-based radiation therapy (RT). METHODS: A pitch-enabled head positioner with an open-face mask were used and, to evaluate inter- and intrafraction variations, 1-3 Cone-Beam Computed Tomography (CBCT) were performed. Surface Guided Radiation Therapy (SGRT) was used to evaluate intrafraction variations at 3 time points: initial (i), final (f), and monitoring (m) (before, end, and during RT). Data regarding technician mask aspect were collected. RESULTS: Between October 2019 and April 2020, 69 patients with brain disease were treated: 45 received stereotactic RT and 24 conventional RT; 556 treatment sessions and 863 CBCT's were performed. Inter-fraction CBCT mean values were longitudinally 0.9 mm, laterally 0.8 mm, vertically 1.1 mm, roll 0.58°, pitch 0.59°, yaw 0.67°. Intrafraction CBCT mean values were longitudinally 0.3 mm, laterally 0.3 mm, vertically 0.4 mm, roll 0.22°, pitch 0.33°, yaw 0.24°. SGRT intrafraction mean values were: i_, m_, f_ longitudinally 0.09 mm, 0.45 mm, 0.31 mm; i_, m_, f_ laterally 0.07 mm, 0.36 mm, 0.20 mm; i_, m_, f_ vertically 0.06 mm, 0.31 mm, 0.22 mm; i_, m_, f_ roll 0.025°, 0.208°, 0.118°; i_, m_, f_ pitch 0.036°, 0.307°, 0.194°; i_, m_, f_ yaw 0.039°, 0.274°, 0.189°. CONCLUSIONS: This immobilization solution is reproducible and stable. Combining CBCT and SGRT data confirm that 1 mm CTV-PTV margin for Linac-based SRT was adequate. Using open-face mask and SGRT, for conventional RT, radiological imaging could be omitted.
RESUMO
BACKGROUND AND OBJECTIVE: The research is done in the field of Augmented Reality (AR) for patient positioning in radiation therapy is scarce. We propose an efficient and cost-effective algorithm for tracking the scene and the patient to interactively assist the patient's positioning process by providing visual feedback to the operator. Up to our knowledge, this is the first framework that can be employed for mobile interactive AR to guide patient positioning. METHODS: We propose a pointcloud processing method that, combined with a fiducial marker-mapper algorithm and the generalized ICP algorithm, tracks the patient and the camera precisely and efficiently only using the CPU unit. The 3D reference model and body marker map alignment is calculated employing an efficient body reconstruction algorithm. RESULTS: Our quantitative evaluation shows that the proposed method achieves a translational and rotational error of 4.17 mm/0.82∘ at 9 fps. Furthermore, the qualitative results demonstrate the usefulness of our algorithm in patient positioning on different human subjects. CONCLUSION: Since our algorithm achieves a relatively high frame rate and accuracy employing a regular laptop (without a dedicated GPU), it is a very cost-effective AR-based patient positioning method. It also opens the way for other researchers by introducing a framework that could be improved upon for better mobile interactive AR patient positioning solutions in the future.