The effect of user-defined variables on dosimetry consistency in Gamma Knife planning.
Ma, L; Chin, L S; Shepard, D; Amin, P; Slawson, R.
Afiliação
  • Ma L; Department of Radiation Oncology, University of Maryland School of Medicine, USA. lma001@umaryland.edu
Phys Med Biol ; 45(5): N43-47, 2000 May.
Article em En | MEDLINE | ID: mdl-10843114
We report a dosimetric variation caused by a user-defined variable for the Leksell Gamma Knife planning system. Treatment plans of 31 randomly selected patients were studied retrospectively to determine the dosimetric effects in the dose prescription and computation as a result of dose matrix positioning in the Leksell Gamma Plan (LGP, Version 4.12). Phantom studies with ion chamber measurements were carried out to validate the accuracy of the computation results. An average overdose of 2% was found due to the variations in the user-defined dose matrix position for the studied cases. In the extreme, the overdose value was as high as 5% with an over-treatment time exceeding 2 min. The phantom measurements were found to agree with the LGP calculation within 0.5%. An adaptive method was developed and demonstrated in this study to eliminate such dosimetry variations.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Radiocirurgia Tipo de estudo: Observational_studies Limite: Humans Idioma: En Ano de publicação: 2000 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Radiocirurgia Tipo de estudo: Observational_studies Limite: Humans Idioma: En Ano de publicação: 2000 Tipo de documento: Article