Your browser doesn't support javascript.
loading
Evaluation of soft tissue response to a poly(urethane urea).
Mohanty, M; Hunt, J A; Doherty, P J; Annis, D; Williams, D F.
Afiliação
  • Mohanty M; Pathophysiology Division, Sree Chitra Tirunal Institute of Medical Sciences and Technology, Trivandrum, India.
Biomaterials ; 13(10): 651-6, 1992.
Article em En | MEDLINE | ID: mdl-1420709
ABSTRACT
Variations in the performance of vascular prostheses constructed of polyurethanes, and some evidence which suggested that these variations could be due not to the properties of the polymer itself, but to differences in the cellular response to the various microstructures of porous polyurethanes require investigation. Experiments were performed to evaluate quantitatively the extent of the cell behaviour adjacent to a series of polyurethane samples. It was shown that, with Biomer, a polyurethane urea, the profile of cell behaviour as a function of distance from the implant surface and of time following implantation, the response of cells in general and macrophages in particular, varied considerably with different internal microstructure. This supports the suggestion that the cellular response to different structures and susceptibility to degradation are related.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Polímeros / Poliuretanos / Prótese Vascular / Músculos Limite: Animals Idioma: En Ano de publicação: 1992 Tipo de documento: Article
Buscar no Google