Absolute calibration of imaging plate for GeV electrons.
Rev Sci Instrum
; 79(6): 066102, 2008 Jun.
Article
em En
| MEDLINE
| ID: mdl-18601438
ABSTRACT
An imaging plate has been used as a useful detector of energetic electrons in laser electron acceleration and laser fusion studies. The absolute sensitivity of an imaging plate was calibrated at 1 GeV electron energy using the injector Linac of SPring-8. The sensitivity curve obtained up to 100 MeV in a previous study was extended successfully to GeV range.