A twelve-analyzer detector system for high-resolution powder diffraction.
Lee, Peter L; Shu, Deming; Ramanathan, Mohan; Preissner, Curt; Wang, Jun; Beno, Mark A; Von Dreele, Robert B; Ribaud, Lynn; Kurtz, Charles; Antao, Sytle M; Jiao, Xuesong; Toby, Brian H.
Afiliação
  • Lee PL; Advanced Photon Source, Argonne National Laboratory, Argonne, IL 60439, USA. pllee@aps.anl.gov
J Synchrotron Radiat ; 15(Pt 5): 427-32, 2008 Sep.
Article em En | MEDLINE | ID: mdl-18728312
ABSTRACT
A dedicated high-resolution high-throughput X-ray powder diffraction beamline has been constructed at the Advanced Photon Source (APS). In order to achieve the goals of both high resolution and high throughput in a powder instrument, a multi-analyzer detector system is required. The design and performance of the 12-analyzer detector system installed on the powder diffractometer at the 11-BM beamline of APS are presented.