A fast double template convolution isocenter evaluation algorithm with subpixel accuracy.
Winey, Brian; Sharp, Greg; Bussière, Marc.
Afiliação
  • Winey B; Harvard Medical School, Massachusetts General Hospital, Boston, Massachusetts 02114, USA. winey.brian@mgh.harvard.edu
Med Phys ; 38(1): 223-7, 2011 Jan.
Article em En | MEDLINE | ID: mdl-21361190
ABSTRACT

PURPOSE:

To design a fast Winston Lutz (fWL) algorithm for accurate analysis of radiation isocenter from images without edge detection or center of mass calculations.

METHODS:

An algorithm has been developed to implement the Winston Lutz test for mechanical/ radiation isocenter agreement using an electronic portal imaging device (EPID). The algorithm detects the position of the radiation shadow of a tungsten ball within a stereotactic cone. The fWL algorithm employs a double convolution to independently find the position of the sphere and cone centers. Subpixel estimation is used to achieve high accuracy. Results of the algorithm were compared to (1) a human observer with template guidance and (2) an edge detection/center of mass (edCOM) algorithm. Testing was performed with high resolution (0.05 mm/px, film) and low resolution (0.78 mm/px, EPID) image sets.

RESULTS:

Sphere and cone center relative positions were calculated with the fWL algorithm for high resolution test images with an accuracy of 0.002 +/- 0.061 mm compared to 0.042 +/- 0.294 mm for the human observer, and 0.003 +/- 0.038 mm for the edCOM algorithm. The fWL algorithm required 0.01 s per image compared to 5 s for the edCOM algorithm and 20 s for the human observer. For lower resolution images the fWL algorithm localized the centers with an accuracy of 0.083 +/- 0.12 mm compared to 0.03 +/- 0.5514 mm for the edCOM algorithm.

CONCLUSIONS:

A fast (subsecond) subpixel algorithm has been developed that can accurately determine the center locations of the ball and cone in Winston Lutz test images without edge detection or COM calculations.