Your browser doesn't support javascript.
loading
Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification--a multi-vendor study.
Rolf, Marijn P; Hofman, Mark B M; Gatehouse, Peter D; Markenroth-Bloch, Karin; Heymans, Martijn W; Ebbers, Tino; Graves, Martin J; Totman, John J; Werner, Beat; van Rossum, Albert C; Kilner, Philip J; Heethaar, Rob M.
Afiliação
  • Rolf MP; Department of Physics and Medical Technology, ICaR-VU, VU University Medical Center, Amsterdam, the Netherlands. mp.rolf@vumc.nl
J Cardiovasc Magn Reson ; 13: 18, 2011 Mar 09.
Article em En | MEDLINE | ID: mdl-21388521
ABSTRACT

PURPOSE:

Eddy current induced velocity offsets are of concern for accuracy in cardiovascular magnetic resonance (CMR) volume flow quantification. However, currently known theoretical aspects of eddy current behavior have not led to effective guidelines for the optimization of flow quantification sequences. This study is aimed at identifying correlations between protocol parameters and the resulting velocity error in clinical CMR flow measurements in a multi-vendor study.

METHODS:

Nine 1.5T scanners of three different types/vendors were studied. Measurements were performed on a large stationary phantom. Starting from a clinical breath-hold flow protocol, several protocol parameters were varied. Acquisitions were made in three clinically relevant orientations. Additionally, a time delay between the bipolar gradient and read-out, asymmetric versus symmetric velocity encoding, and gradient amplitude and slew rate were studied in adapted sequences as exploratory measurements beyond the protocol. Image analysis determined the worst-case offset for a typical great-vessel flow measurement.

RESULTS:

The results showed a great variation in offset behavior among scanners (standard deviation among samples of 0.3, 0.4, and 0.9 cm/s for the three different scanner types), even for small changes in the protocol. Considering the absolute values, none of the tested protocol settings consistently reduced the velocity offsets below the critical level of 0.6 cm/s neither for all three orientations nor for all three scanner types. Using multilevel linear model analysis, oblique aortic and pulmonary slices showed systematic higher offsets than the transverse aortic slices (oblique aortic 0.6 cm/s, and pulmonary 1.8 cm/s higher than transverse aortic). The exploratory measurements beyond the protocol yielded some new leads for further sequence development towards reduction of velocity offsets; however those protocols were not always compatible with the time-constraints of breath-hold imaging and flow-related artefacts.

CONCLUSIONS:

This study showed that with current systems there was no generic protocol which resulted into acceptable flow offset values. Protocol optimization would have to be performed on a per scanner and per protocol basis. Proper optimization might make accurate (transverse) aortic flow quantification possible for most scanners. Pulmonary flow quantification would still need further (offline) correction.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aorta / Imageamento por Ressonância Magnética / Circulação Pulmonar Tipo de estudo: Clinical_trials / Guideline / Prognostic_studies Limite: Humans País/Região como assunto: Europa Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aorta / Imageamento por Ressonância Magnética / Circulação Pulmonar Tipo de estudo: Clinical_trials / Guideline / Prognostic_studies Limite: Humans País/Região como assunto: Europa Idioma: En Ano de publicação: 2011 Tipo de documento: Article