ABSTRACT
The gamma reaction history diagnostic at the National Ignition Facility has the capability to determine a number of important performance metrics for cryogenic deuterium-tritium implosions the fusion burn width, bang time and yield, as well as the areal density of the compressed ablator. Extracting those values from the measured γ rays of an implosion, requires accounting for a γ-ray background in addition to the impulse response function of the instrument. To address these complications, we have constructed a model of the γ-ray signal, and are developing a simultaneous multi-shot fitting routine to constrain its parameter space.

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Ano de publicação: 2012 Tipo de documento: Article