T1 mapping for myocardial extracellular volume measurement by CMR: bolus only versus primed infusion technique.
White, Steven K; Sado, Daniel M; Fontana, Marianna; Banypersad, Sanjay M; Maestrini, Viviana; Flett, Andrew S; Piechnik, Stefan K; Robson, Matthew D; Hausenloy, Derek J; Sheikh, Amir M; Hawkins, Philip N; Moon, James C.
Afiliação
  • White SK; The Heart Hospital, London, United Kingdom; Institute of Cardiovascular Science, University College London, London, United Kingdom; The Hatter Cardiovascular Institute, University College London, London, United Kingdom.
JACC Cardiovasc Imaging ; 6(9): 955-62, 2013 Sep.
Article em En | MEDLINE | ID: mdl-23582361
ABSTRACT

OBJECTIVES:

The aim of this study was to determine the accuracy of the contrast "bolus only" T1 mapping cardiac magnetic resonance (CMR) technique for measuring myocardial extracellular volume fraction (ECV).

BACKGROUND:

Myocardial ECV can be measured with T1 mapping before and after contrast agent if the contrast agent distribution between blood/myocardium is at equilibrium. Equilibrium distribution can be achieved with a primed contrast infusion (equilibrium contrast-CMR [EQ-CMR]) or might be approximated by the dynamic equilibration achieved by delayed post-bolus measurement. This bolus only approach is highly attractive, but currently limited data support its use. We compared the bolus only technique with 2 independent standards collagen volume fraction (CVF) from myocardial biopsy in aortic stenosis (AS); and the infusion technique in 5 representative conditions.

METHODS:

One hundred forty-seven subjects were studied healthy volunteers (n = 50); hypertrophic cardiomyopathy (n = 25); severe AS (n = 22); amyloid (n = 20); and chronic myocardial infarction (n = 30). Bolus only (at 15 min) and infusion ECV measurements were performed and compared. In 18 subjects with severe AS the results were compared with histological CVF.

RESULTS:

The ECV by both techniques correlated with histological CVF (n = 18, r² = 0.69, p < 0.01 vs. r² = 0.71, p < 0.01, p = 0.42 for comparison). Across health and disease, there was strong correlation between the techniques (r² = 0.97). However, in diseases of high ECV (amyloid, hypertrophic cardiomyopathy late gadolinium enhancement, and infarction), Bland-Altman analysis indicates the bolus only technique has a consistent and increasing offset, giving a higher value for ECVs above 0.4 (mean difference ± limit of agreement for ECV <0.4 = -0.004 ± 0.037 vs. ECV >0.4 = 0.040 ± 0.075, p < 0.001).

CONCLUSIONS:

Bolus only, T1 mapping-derived ECV measurement is sufficient for ECV measurement across a range of cardiac diseases, and this approach is histologically validated in AS. However, when ECV is >0.4, the bolus only technique consistently measures ECV higher compared with infusion.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cardiomiopatia Hipertrófica / Imagem Cinética por Ressonância Magnética / Matriz Extracelular / Miocárdio Tipo de estudo: Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cardiomiopatia Hipertrófica / Imagem Cinética por Ressonância Magnética / Matriz Extracelular / Miocárdio Tipo de estudo: Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article