An improved cation exchange HPLC method for the measurement of serum creatinine.
Guy, J M; Legg, E F.
Afiliação
  • Guy JM; Department of Clinical Chemistry, East Birmingham Hospital, Bordesley Green East, UK.
Ann Clin Biochem ; 27 ( Pt 3): 223-6, 1990 May.
Article em En | MEDLINE | ID: mdl-2382956
ABSTRACT
An improved HPLC method for the measurement of serum creatinine is described. Separation is effected using a strong cation exchange column at pH 4.7. Analytical recovery, precision and linearity are satisfactory, and the method correlates well with a kinetic Jaffé procedure. The method provides a means of accurately measuring creatinine and may be of use in the investigation of interference in creatinine assays.