State of the art for measurement of urine albumin: comparison of routine measurement procedures to isotope dilution tandem mass spectrometry.
Bachmann, Lorin M; Nilsson, Goran; Bruns, David E; McQueen, Matthew J; Lieske, John C; Zakowski, Jack J; Miller, W Greg.
Afiliação
  • Bachmann LM; Department of Pathology, Virginia Commonwealth University, Richmond, VA;
Clin Chem ; 60(3): 471-80, 2014 Mar.
Article em En | MEDLINE | ID: mdl-24281781
BACKGROUND: Urine albumin is the primary biomarker for detection and monitoring of kidney damage. Because fixed decision criteria are used to identify patients with increased values, we investigated if commonly used routine measurement procedures gave comparable results. METHODS: Results from 17 commercially available urine albumin measurement procedures were investigated vs an isotope dilution mass spectrometry (IDMS) procedure. Nonfrozen aliquots of freshly collected urine from 332 patients with chronic kidney disease, diabetes, cardiovascular disease, and hypertension were distributed to manufacturers to perform urine albumin measurements according to the respective instructions for use for each procedure. Frozen aliquots were used for measurements by the IDMS procedure. An error model was used to determine imprecision and bias components. RESULTS: Median differences between the largest positive and negative biases vs IDMS were 45%, 37%, and 42% in the concentration intervals of 12-30 mg/L, 31-200 mg/L, and 201-1064 mg/L, respectively. Biases varied with concentration for most procedures and exceeded ± 10% over the concentration interval for 14 of 16 quantitative procedures. Mean biases ranged from -35% to 34% at 15 mg/L. Dilution of samples with high concentrations introduced bias for 4 procedures. The combined CV was >10% for 5 procedures. It was not possible to estimate total error due to dependence of bias on concentration. CVs for sample-specific influences were 0% to 15.2%. CONCLUSIONS: Bias was the dominant source of disagreement among routine measurement procedures. Consequently, standardization efforts will improve agreement among results. Variation of bias with concentration needs to be addressed by manufacturers.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Albumina Sérica Tipo de estudo: Prognostic_studies Limite: Adult / Aged / Aged80 / Humans / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Albumina Sérica Tipo de estudo: Prognostic_studies Limite: Adult / Aged / Aged80 / Humans / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article