Your browser doesn't support javascript.
loading
Advances in real-time phase-contrast flow MRI using asymmetric radial gradient echoes.
Untenberger, Markus; Tan, Zhengguo; Voit, Dirk; Joseph, Arun A; Roeloffs, Volkert; Merboldt, K Dietmar; Schätz, Sebastian; Frahm, Jens.
Afiliação
  • Untenberger M; Biomedizinische NMR Forschungs GmbH am Max-Planck-Institut für biophysikalische Chemie, Göttingen, Germany.
  • Tan Z; Biomedizinische NMR Forschungs GmbH am Max-Planck-Institut für biophysikalische Chemie, Göttingen, Germany.
  • Voit D; Biomedizinische NMR Forschungs GmbH am Max-Planck-Institut für biophysikalische Chemie, Göttingen, Germany.
  • Joseph AA; Biomedizinische NMR Forschungs GmbH am Max-Planck-Institut für biophysikalische Chemie, Göttingen, Germany.
  • Roeloffs V; DZHK, German Center for Cardiovascular Research, partner site Göttingen, Germany.
  • Merboldt KD; Biomedizinische NMR Forschungs GmbH am Max-Planck-Institut für biophysikalische Chemie, Göttingen, Germany.
  • Schätz S; Biomedizinische NMR Forschungs GmbH am Max-Planck-Institut für biophysikalische Chemie, Göttingen, Germany.
  • Frahm J; Biomedizinische NMR Forschungs GmbH am Max-Planck-Institut für biophysikalische Chemie, Göttingen, Germany.
Magn Reson Med ; 75(5): 1901-8, 2016 May.
Article em En | MEDLINE | ID: mdl-26096085
ABSTRACT

PURPOSE:

To provide multidimensional velocity compensation for real-time phase-contrast flow MRI.

METHODS:

The proposed method introduces asymmetric gradient echoes for highly undersampled radial FLASH MRI with phase-sensitive image reconstruction by regularized nonlinear inversion (NLINV). Using an adapted gradient delay correction the resulting image quality was analyzed by simulations and experimentally validated at 3 Tesla. For real-time flow MRI the reduced gradient-echo timing allowed for the incorporation of velocity-compensating waveforms for all imaging gradients at even shorter repetition times.

RESULTS:

The results reveal a usable degree of 20% asymmetry. Real-time flow MRI with full velocity compensation eliminated signal void in a flow phantom, confirmed flow parameters in healthy subjects and demonstrated signal recovery and phase conservation in a patient with aortic valve insufficiency and stenosis. Exemplary protocols at 1.4-1.5 mm resolution and 6 mm slice thickness achieved total acquisition times of 33.3-35.7 ms for two images (7 spokes each) with and without flow-encoding gradient.

CONCLUSION:

Asymmetric gradient echoes were successfully implemented for highly undersampled radial trajectories. The resulting temporal gain offers full velocity compensation for real-time phase-contrast flow MRI which minimizes false-positive contributions from complex flow and further enhances the temporal resolution compared with acquisitions with symmetric echoes.