Intra-Aortic Balloon Pump Malposition Reduces Visceral Artery Perfusion in an Acute Animal Model.
Vondran, Maximilian; Rastan, Ardawan J; Tillmann, Eugen; Seeburger, Jörg; Schröter, Thomas; Dhein, Stefan; Bakhtiary, Farhad; Mohr, Friedrich-Wilhelm.
Afiliação
  • Vondran M; Department of Cardiac Surgery, Heart Center Leipzig, University of Leipzig, Leipzig, Germany.
  • Rastan AJ; Department of Cardiac Surgery, Heart Center Leipzig, University of Leipzig, Leipzig, Germany.
  • Tillmann E; Department of Cardiac Surgery, Heart Center Leipzig, University of Leipzig, Leipzig, Germany.
  • Seeburger J; Department of Cardiac Surgery, Heart Center Leipzig, University of Leipzig, Leipzig, Germany.
  • Schröter T; Department of Cardiac Surgery, Heart Center Leipzig, University of Leipzig, Leipzig, Germany.
  • Dhein S; Department of Cardiac Surgery, Heart Center Leipzig, University of Leipzig, Leipzig, Germany.
  • Bakhtiary F; Department of Cardiac Surgery, Heart Center Leipzig, University of Leipzig, Leipzig, Germany.
  • Mohr FW; Department of Cardiac Surgery, Heart Center Leipzig, University of Leipzig, Leipzig, Germany.
Artif Organs ; 40(4): 334-40, 2016 Apr.
Article em En | MEDLINE | ID: mdl-26366459
ABSTRACT
Visceral artery perfusion can be potentially affected by intra-aortic balloon pump (IABP) catheters. We utilized an animal model to quantify the acute impact of a low balloon position on mesenteric artery perfusion. In six pigs (78 ± 7 kg), a 30-cc IABP was placed in the descending aorta in a transfemoral procedure. The celiac artery (CA) and the cranial mesenteric artery (CMA) were surgically dissected. Transit time blood flow was measured for (i) baseline, (ii) 11 augmentation with the balloon proximal to the visceral arteries, and (iii) 11 augmentation with the balloon covering the visceral arteries. Blood flow in the CMA and CA was reduced by 17 and 24%, respectively, when the balloon compromised visceral arteries compared with a position above the visceral arteries (flow in mL/min CMA (i) 1281 ± 512, (ii) 1389 ± 287, (iii) 1064 ± 276, P < 0.05 for 3 vs. 1 and 3 vs. 2; CA (i) 885 ± 370, (ii) 819 ± 297, (iii) 673 ± 315; P < 0.05 for 3 vs. 1). The covering of visceral arteries by an IABP balloon causes a significant reduction of visceral artery perfusion; thus, the positioning of this device during implantation is critical for obtaining a satisfactory outcome.