A field-deployed surface plasmon resonance (SPR) sensor for RDX quantification in environmental waters.
Brulé, Thibault; Granger, Geneviève; Bukar, Natalia; Deschênes-Rancourt, Clarisse; Havard, Thierry; Schmitzer, Andreea R; Martel, Richard; Masson, Jean-Francois.
Afiliação
  • Brulé T; Department of Chemistry, Université de Montréal, C.P. 6128 Succ. Centre-Ville, Montreal, Quebec, H3C 3J7 Canada. jf.masson@umontreal.ca.
  • Granger G; Department of Chemistry, Université de Montréal, C.P. 6128 Succ. Centre-Ville, Montreal, Quebec, H3C 3J7 Canada. jf.masson@umontreal.ca.
  • Bukar N; Department of Chemistry, Université de Montréal, C.P. 6128 Succ. Centre-Ville, Montreal, Quebec, H3C 3J7 Canada. jf.masson@umontreal.ca.
  • Deschênes-Rancourt C; Institut national de la recherche scientifique, Eau Terre Environnement Centre (ETE-INRS), 490 rue de la Couronne, Quebec city, G1K 9A9, Canada.
  • Havard T; Department of Chemistry, Université de Montréal, C.P. 6128 Succ. Centre-Ville, Montreal, Quebec, H3C 3J7 Canada. jf.masson@umontreal.ca.
  • Schmitzer AR; Department of Chemistry, Université de Montréal, C.P. 6128 Succ. Centre-Ville, Montreal, Quebec, H3C 3J7 Canada. jf.masson@umontreal.ca.
  • Martel R; Institut national de la recherche scientifique, Eau Terre Environnement Centre (ETE-INRS), 490 rue de la Couronne, Quebec city, G1K 9A9, Canada.
  • Masson JF; Department of Chemistry, Université de Montréal, C.P. 6128 Succ. Centre-Ville, Montreal, Quebec, H3C 3J7 Canada. jf.masson@umontreal.ca and Centre for Self-Assembled Chemical Structures (CSACS), 801 Sherbrooke St. West, Montreal, Quebec, H3A 2K6 Canada.
Analyst ; 142(12): 2161-2168, 2017 Jun 21.
Article em En | MEDLINE | ID: mdl-28548156
ABSTRACT
A field-deployable surface plasmon resonance (SPR) sensor is reported for the detection of the energetic material (commonly termed explosives) 1,3,5-trinitroperhydro-1,3,5-triazine (RDX) at ppb concentration in environmental samples. The SPR sensor was first validated under laboratory conditions with uncontaminated natural water samples spiked with known concentrations of RDX near the EPA limit of 2 ppb, which was then applied to monitor environmental samples collected in different downgradient wells near a grenade training range. The SPR sensor was finally tested on the field, where environmental samples were analysed on location in less than 90 minutes per well, which included the time to setup the equipment, sample the well and analyse the sample. The SPR analysis time was less than 45 minutes for equilibration, recalibration and measuring the water sample. Results obtained with the SPR sensors were cross-validated with the standard HPLC method (EPA method 8330b), and they showed good agreement with an accuracy within less than 1.6 ppb for analysis at the sampling sites, and with the relative standard deviation (RSD) better than 20% for field and laboratory measurements. The SPR sensor worked in a range of environmental conditions, including operation from about 0 °C to nearly 30 °C. The instrument was easily deployed near the sampling site using motor vehicles under summer conditions (Lab-in-a-Jeep) and using a sled under winter conditions (Lab-on-a-sled), showcasing the field deployability of the RDX SPR sensor and the possibility of continuously monitoring RDX in the environment.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2017 Tipo de documento: Article