Your browser doesn't support javascript.
loading
Adaption of a PIN-diode detector as an online neutron monitor for the thermal column of the TRIGA research reactor.
Schmitz, Tobias; Blaickner, Matthias.
Afiliação
  • Schmitz T; Institut für Kernchemie, Johannes Gutenberg-Universität, Mainz DE-55128, Germany.
  • Blaickner M; Center of Health & Bioresources, AIT Austrian Institute of Technology, Donau-City-Strasse 1/2, A-1220 Vienna, Austria. Electronic address: matthias.blaickner@ait.ac.at.
Appl Radiat Isot ; 128: 142-147, 2017 Oct.
Article em En | MEDLINE | ID: mdl-28710934
A BNCT online neutron monitoring system was tested in a TRIGA reactor, using a silicon PIN-diode with a conversion foil. The setup was tested with different reactor powers at the hot and cold ends of the irradiation channel, using activation foils to compare with measured fluxes. The results demonstrate good reproducibility and show a linear correlation between signal of the PIN-diode and neutron flux at all positions, demonstrating this approach to be suitable for online monitoring of the neutron flux.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2017 Tipo de documento: Article