Should Single-Coil Implantable Cardioverter Defibrillator Leads Be Used in all Patients?
Card Electrophysiol Clin
; 10(1): 59-66, 2018 03.
Article
em En
| MEDLINE
| ID: mdl-29428142
ABSTRACT
The historical preference for dual-coil implantable cardioverter defibrillator leads stems from high defibrillation thresholds associated with old device platforms. The high safety margins generated by contemporary devices have rendered the modest difference in defibrillation efficacy between single- and dual-coil leads clinically insignificant. Cohort data demonstrating worse lead extraction outcomes and higher all-cause mortality have brought the incremental utility of an superior vena cava coil into question. This article summarizes the current literature and re-evaluates the utility of dual-coil leads in the context of modern device technology.