Experimental assessment of a 3-D plenoptic endoscopic imaging system.
Le, Hanh N D; Decker, Ryan; Krieger, Axel; Kang, Jin U.
Afiliação
  • Le HND; Electrical and Computer Engineering Department, Johns Hopkins University, Baltimore, MD 21218, USA.
  • Decker R; Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Health System, Washington, DC 20010, USA.
  • Krieger A; Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Health System, Washington, DC 20010, USA.
  • Kang JU; Electrical and Computer Engineering Department, Johns Hopkins University, Baltimore, MD 21218, USA.
Chin Opt Lett ; 15(5)2017.
Article em En | MEDLINE | ID: mdl-29449863
ABSTRACT
An endoscopic imaging system using a plenoptic technique to reconstruct 3-D information is demonstrated and analyzed in this Letter. The proposed setup integrates a clinical surgical endoscope with a plenoptic camera to achieve a depth accuracy error of about 1 mm and a precision error of about 2 mm, within a 25 mm × 25 mm field of view, operating at 11 frames per second.