Non-Invasive Electromagnetic Skin Patch Sensor to Measure Intracranial Fluid-Volume Shifts.
Griffith, Jacob; Cluff, Kim; Eckerman, Brandon; Aldrich, Jessica; Becker, Ryan; Moore-Jansen, Peer; Patterson, Jeremy.
Afiliação
  • Griffith J; Biomedical Engineering, Wichita State University, Wichita, KS 67260, USA. jlgriffith2@shockers.wichita.edu.
  • Cluff K; Biomedical Engineering, Wichita State University, Wichita, KS 67260, USA. kim.cluff@wichita.edu.
  • Eckerman B; Biomedical Engineering, Wichita State University, Wichita, KS 67260, USA. bdeckerman@shockers.wichita.edu.
  • Aldrich J; Biomedical Engineering, Wichita State University, Wichita, KS 67260, USA. jxaldrich@shockers.wichita.edu.
  • Becker R; Biomedical Engineering, Wichita State University, Wichita, KS 67260, USA. linkxor@gmail.com.
  • Moore-Jansen P; Department of Anthropology, Wichita State University, Wichita, KS 67260, USA. pmojan@wichita.edu.
  • Patterson J; Human Performance Studies, Wichita State University, Wichita, KS 67260, USA. jeremy.patterson@wichita.edu.
Sensors (Basel) ; 18(4)2018 Mar 29.
Article em En | MEDLINE | ID: mdl-29596338
ABSTRACT
Elevated intracranial fluid volume can drive intracranial pressure increases, which can potentially result in numerous neurological complications or death. This study's focus was to develop a passive skin patch sensor for the head that would non-invasively measure cranial fluid volume shifts. The sensor consists of a single baseline component configured into a rectangular planar spiral with a self-resonant frequency response when impinged upon by external radio frequency sweeps. Fluid volume changes (10 mL increments) were detected through cranial bone using the sensor on a dry human skull model. Preliminary human tests utilized two sensors to determine feasibility of detecting fluid volume shifts in the complex environment of the human body. The correlation between fluid volume changes and shifts in the first resonance frequency using the dry human skull was classified as a second order polynomial with R² = 0.97. During preliminary and secondary human tests, a ≈24 MHz and an average of ≈45.07 MHz shifts in the principal resonant frequency were measured respectively, corresponding to the induced cephalad bio-fluid shifts. This electromagnetic resonant sensor may provide a non-invasive method to monitor shifts in fluid volume and assist with medical scenarios including stroke, cerebral hemorrhage, concussion, or monitoring intracranial pressure.