Pupil mask diversity for image correction in microscopy.
Opt Express ; 26(12): 14832-14841, 2018 Jun 11.
Article em En | MEDLINE | ID: mdl-30114789
ABSTRACT
Three-dimensional microscopy suffers from sample-induced aberrations that reduce the resolution and lead to misinterpretations of the object distribution. In this paper, the resolution of a three-dimensional fluorescent microscope is significantly improved by introducing an amplitude diversity in the form of a binary amplitude mask positioned in several different orientations within the pupil, followed by computer processing of the diversity images. The method has proved to be fast, easy to implement, and cost-effective in high-resolution imaging of casper fliGFP zebrafish.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2018 Tipo de documento: Article