Your browser doesn't support javascript.
loading
Modeling AeroForm tissue expander for postmastectomy radiation therapy.
Dziemianowicz, Elaine; Gardner, Stephen J; Chin Snyder, Karen; Wen, Ning; Walker, Eleanor M; Fraser, Correen; Reding, Anne; Chetty, Indrin J.
Afiliação
  • Dziemianowicz E; Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
  • Gardner SJ; Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
  • Chin Snyder K; Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
  • Wen N; Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
  • Walker EM; Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
  • Fraser C; Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
  • Reding A; Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
  • Chetty IJ; Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
J Appl Clin Med Phys ; 20(8): 87-97, 2019 Aug.
Article em En | MEDLINE | ID: mdl-31332943
ABSTRACT
The AeroForm chest wall tissue expander (TE) is a silicon shell containing a metallic CO2 reservoir, placed surgically after mastectomy. The patient uses a remote control to release compressed CO2 from the reservoir to inflate the expander. AeroForm poses challenges in a radiation therapy

setting:

The high density of the metallic reservoir causes imaging artifacts on the planning CT, which encumber structure definition and cause misrepresentation of density information, in turn affecting dose calculation. Additionally, convolution-based dose calculation algorithms may not be well-suited to calculate dose in and around high-density materials. In this study, a model of the AeroForm TE was created in Eclipse treatment planning system (TPS). The TPS model was validated by comparing measured to calculated transmission through the AeroForm. Transmission was measured with various geometries using radiochromic film. Dose was calculated with both Varian's Anisotropic Analytical Algorithm (AAA) and Acuros External Beam (AXB) algorithms. AAA and AXB were compared using dose profile and gamma analyses. While both algorithms modeled direct transmission well, AXB better modeled lateral scatter from the AeroForm TE. Clinical significance was evaluated using clinical data from four patients with AeroForm TEs. The AeroForm TPS model was applied, and RT plans were optimized using AAA, then re-calculated with AXB. Structures of clinical significance were defined and dose volume histogram analysis was performed. Compared to AXB, AAA overestimates dose in the AeroForm device. Changes in clinically significant regions were patient- and plan-specific. This study proposes a clinical procedure for modeling the AeroForm in a commercial TPS, and discusses the limitations of dose calculation in and around the device. An understanding of dose calculation accuracy in the vicinity of the AeroForm is critical for assessing individual plan quality, appropriateness of different planning techniques and dose calculation algorithms, and even the decision to use the AeroForm in a postmastectomy radiation therapy setting.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Simulação por Computador / Planejamento da Radioterapia Assistida por Computador / Neoplasias da Mama / Dispositivos para Expansão de Tecidos / Imagens de Fantasmas / Mastectomia / Modelos Teóricos Tipo de estudo: Observational_studies / Prognostic_studies Limite: Female / Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Simulação por Computador / Planejamento da Radioterapia Assistida por Computador / Neoplasias da Mama / Dispositivos para Expansão de Tecidos / Imagens de Fantasmas / Mastectomia / Modelos Teóricos Tipo de estudo: Observational_studies / Prognostic_studies Limite: Female / Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article