Design of ionophores for ion-selective microsensors.
Can J Physiol Pharmacol ; 65(5): 879-84, 1987 May.
Article em En | MEDLINE | ID: mdl-3304589
Requirements for a reliable use of liquid membrane microelectrodes are discussed in terms of stability, response time, and lifetime on the basis of membrane technological considerations. The selectivity of H+, Li+, Na+, K+, Mg2+, Ca2+, and Cl- microelectrodes is critically evaluated using the Nikolskii-Eisenman formalism. Recent progress in the design of new ionophores is presented. A novel neutral carrier-based Ca2+-selective microelectrode with a detection limit of about 5 X 10(-10) M Ca2+ at a background of 125 mM K+ has been realized. An neutral carrier-based microelectrode for H+ with extended pH range of the sample solution is now available. Promising developments in the field of Li+-, Mg2+-, and Cl--selective ionophores are discussed.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Eletroquímica / Ionóforos / Íons Limite: Animals Idioma: En Ano de publicação: 1987 Tipo de documento: Article