Performance characteristics of five antigen-detecting rapid diagnostic test (Ag-RDT) for SARS-CoV-2 asymptomatic infection: a head-to-head benchmark comparison.
J Infect
; 82(6): 269-275, 2021 Jun.
Article
em En
| MEDLINE
| ID: mdl-33882299
BACKGROUND: Mass testing for early identification and isolation of infectious COVID-19 individuals is efficacious for reducing disease spread. Antigen-detecting rapid diagnostic tests (Ag-RDT) may be suitable for testing strategies; however, benchmark comparisons are scarce. METHODS: We used 286 nasopharyngeal specimens from unexposed asymptomatic individuals collected between December 2020 and January 2021 to assess five Ag-RDTs marketed by Abbott, Siemens, Roche Diagnostics, Lepu Medical, and Surescreen. RESULTS: For the overall sample, the performance parameters of Ag-RDTs were as follows: Abbott assay, sensitivity 38.6% (95%CI 29.1-48.8) and specificity 99.5% (97-100%); Siemens, sensitivity 51.5% (41.3-61.6) and specificity 98.4% (95.3-99.6); Roche, sensitivity 43.6% (33.7-53.8) and specificity 96.2% (92.4-98.5); Lepu, sensitivity 45.5% (35.6-55.8) and specificity 89.2% (83.8-93.3%); Surescreen, sensitivity 28.8% (20.2-38.6) and specificity 97.8% (94.5-99.4%). For specimens with cycle threshold (Ct) <30 in RT-qPCR, all Ag-RDT achieved a sensitivity ≥70%. The modelled negative- and positive-predictive value for 1% prevalence were >99% and <50%, respectively. CONCLUSIONS: When screening unexposed asymptomatic individuals, two Ag-RDTs achieved sensitivity ≥80% for specimens with Ct<30 and specificity ≥96%. The estimated negative predictive value suggests the suitability of Ag-RDTs for mass screenings of SARS-CoV-2 infection in the general population.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
SARS-CoV-2
/
COVID-19
Tipo de estudo:
Diagnostic_studies
/
Prognostic_studies
/
Risk_factors_studies
/
Screening_studies
Limite:
Humans
Idioma:
En
Ano de publicação:
2021
Tipo de documento:
Article