Your browser doesn't support javascript.
loading
Liquid-Immersion Reprocessing Effects on Filtration Efficiency of 'Single-Use' Commercial Medical Face Masks.
Everts, Richard J; Al Ghusaini, Shadha; Telfar-Barnard, Lucy; Barclay, Ella; Tan, Shaun; Jekel, Sonja; Jennings, Lance; Choi, Dong Hoon; Hilson, Dougal; Gibson, Barbara.
Afiliação
  • Everts RJ; Infectious Disease Service, Nelson Bays Primary Health, 281 Queen St, Richmond, Nelson, New Zealand.
  • Al Ghusaini S; University of Otago, 2 Riccarton Ave, Christchurch, New Zealand.
  • Telfar-Barnard L; Department of Public Health, University of Otago, 23a Mein St, Newton, Wellington, New Zealand.
  • Barclay E; University of Otago, 2 Riccarton Ave, Christchurch, New Zealand.
  • Tan S; Lanaco, 2-4 Sultan St, Ellerslie, Auckland, New Zealand.
  • Jekel S; Lanaco, 2-4 Sultan St, Ellerslie, Auckland, New Zealand.
  • Jennings L; Department of Pathology and Biomedical Sciences, University of Otago, 2 Riccarton Ave, Christchurch, New Zealand.
  • Choi DH; Canterbury Health Laboratories, Hagley Ave, Christchurch, New Zealand.
  • Hilson D; Biomedical Engineering Department, Nelson Hospital, Tipahi St, Nelson, New Zealand.
  • Gibson B; University of Otago, 2 Riccarton Ave, Christchurch, New Zealand.
Ann Work Expo Health ; 66(2): 246-259, 2022 02 18.
Article em En | MEDLINE | ID: mdl-34564717
ABSTRACT

PURPOSE:

Medical masks have inferior filtration efficiency and fit to filtering facepiece respirators (FFRs) but are widely used in healthcare and the community. These masks are intended for disposal after use but in the event of mask shortage re-use after reprocessing may be an option. We investigated eight reprocessing methods that each involved washing or soaking in liquid, are likely to eliminate respiratory viruses, and are safe and available in most community and healthcare settings.

METHODS:

Three brands of EN 14683 standards-compliant commercial medical mask were each reprocessed 10 times by one of eight methods. We measured filtration efficiency for poly-dispersed sodium chloride particles and pressure differential.

RESULTS:

Compared with new medical masks, reprocessed masks had significantly reduced filtration efficiency. The reduction was mild-moderate (6.5-25.8%) after warm water wash, hot water soak or boiling water soak; and moderate-large (24.1-51.5%) after detergent, soap or laundry machine wash, or bleach soak. There were mixed and minor changes in pressure differential. Most reprocessed standards-compliant masks had better filtration efficiency than new non-standard commercial masks and then cotton and cotton-polyester mix fabric samples, even triple-layered fabrics.

CONCLUSIONS:

High-quality commercial medical masks reprocessed 10 times by water immersion methods had better filtration efficiency than new non-standard masks and washable fabrics. These findings have particular relevance for community and low-resource healthcare settings.