Balloon Rupture During Balloon Mitral Valvotomy.
Gupta, Preeti; Gupta, Anunay; Agstam, Sourabh; Bansal, Sandeep.
Afiliação
  • Agstam S; VMMC and Safdarjung Hospital, New Delhi, India. sourabhagstam@gmail.com.
J Invasive Cardiol ; 33(11): E916, 2021 Nov.
Article em En | MEDLINE | ID: mdl-34735355
ABSTRACT
The reuse of sterilized Inoue catheters is practiced widely in developing countries to bring down the procedure cost. However, blood can enter the space between the latex layers and become embedded in the mesh layer, which is difficult to clean when sterilizing the catheters. This is a common cause of rupture. Proper meticulous inspection of reused Inoue balloons for deformity or leakage through the small holes is necessary to prevent such complications.
Assuntos
Palavras-chave
Buscar no Google
Base de dados: MEDLINE Assunto principal: Procedimentos Cirúrgicos Cardíacos / Estenose da Valva Mitral Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Procedimentos Cirúrgicos Cardíacos / Estenose da Valva Mitral Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article