Performance of the eazyplex® BloodScreen GN as a simple and rapid molecular test for identification of Gram-negative bacteria from positive blood cultures.
Eur J Clin Microbiol Infect Dis
; 41(3): 489-494, 2022 Mar.
Article
em En
| MEDLINE
| ID: mdl-34807364
The LAMP-based eazyplex® BloodScreen GN was evaluated for the detection of frequent Gram-negatives directly from positive blood culture (BC) bottles. A total of 449 BCs were analyzed. Sensitivities and specificities were 100% and 100% for Escherichia coli, 95.7% and 100% for Klebsiella pneumoniae, 100% and 100% for blaCTX-M, 100% and 100% for Klebsiella oxytoca, 100% and 99% for Proteus mirabilis, and 100% and 99.8% for Pseudomonas aeruginosa, respectively. The time to result ranged from 8 to 16 min, plus about 6 min for sample preparation. The eazyplex® BloodScreen GN is a reliable molecular assay for rapid BC testing.