Your browser doesn't support javascript.
loading
Microdosimetric characterization of a clinical proton therapy beam: comparison between simulated lineal energy distributions in spherical water targets and experimental measurements with a silicon detector.
Parisi, Alessio; Olko, Pawel; Swakon, Jan; Horwacik, Tomasz; Jablonski, Hubert; Malinowski, Leszek; Nowak, Tomasz; Struelens, Lara; Vanhavere, Filip.
Afiliação
  • Parisi A; Belgian Nuclear Research Centre (SCK CEN), Boeretang 200, Mol, Belgium.
  • Olko P; Institute of Nuclear Physics Polish Academy of Sciences (IFJ PAN), Krakow, Poland.
  • Swakon J; Institute of Nuclear Physics Polish Academy of Sciences (IFJ PAN), Krakow, Poland.
  • Horwacik T; Institute of Nuclear Physics Polish Academy of Sciences (IFJ PAN), Krakow, Poland.
  • Jablonski H; Institute of Nuclear Physics Polish Academy of Sciences (IFJ PAN), Krakow, Poland.
  • Malinowski L; Institute of Nuclear Physics Polish Academy of Sciences (IFJ PAN), Krakow, Poland.
  • Nowak T; Institute of Nuclear Physics Polish Academy of Sciences (IFJ PAN), Krakow, Poland.
  • Struelens L; Belgian Nuclear Research Centre (SCK CEN), Boeretang 200, Mol, Belgium.
  • Vanhavere F; Belgian Nuclear Research Centre (SCK CEN), Boeretang 200, Mol, Belgium.
Phys Med Biol ; 67(1)2022 01 17.
Article em En | MEDLINE | ID: mdl-34933289
Objective. Treatment planning based on computer simulations wasproposed to account for the increased relative biological effectiveness (RBE) of proton radiotherapy beams near to the edges of the irradiated volume. Since silicon detectors could be used to validate the results of these simulations, it is important to explore the limitations of this comparison.Approach. Microdosimetric measurements with a MicroPlus Bridge V2 silicon detector (thickness = 10µm) were performed along the Bragg peak of a clinical proton beam. The lineal energy distributions, the dose-mean values, and the RBE calculated with a biological weighting function were compared with PHITS simulations (microdosimetric target = 1µm water sphere), and published clonogenic survivalin vitroRBE data for the V79 cell line. The effect of the silicon-to-water conversion was also investigated by comparing three different methodologies (conversion based on a single value, novel bin-to-bin conversions based on SRIM and PSTAR).Main results. Mainly due to differences in the microdosimetric targets, the experimental dose-mean lineal energy and RBE values at the distal edge were respectively up to 53% and 28% lower than the simulated ones. Furthermore, the methodology chosen for the silicon-to-water conversion was proven to affect the dose-mean lineal energy and the RBE10up to 32% and 11% respectively. The best methodology to compensate for this underestimation was the bin-to-bin silicon-to-water conversion based on PSTAR.Significance. This work represents the first comparison between PHITS-simulated lineal energy distributions in water targets and corresponding experimental spectra measured with silicon detectors. Furthermore, the effect of the silicon-to-water conversion on the RBE was explored for the first time. The proposed methodology based on the PSTAR bin-to-bin conversion appears to provide superior results with respect to commonly used single scaling factors and is recommended for future studies.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Terapia com Prótons Tipo de estudo: Health_economic_evaluation Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Terapia com Prótons Tipo de estudo: Health_economic_evaluation Idioma: En Ano de publicação: 2022 Tipo de documento: Article