Telemonitoring-guided ambulatory fixed CPAP titration versus ambulatory APAP titration in moderate obstructive sleep apnea: A non-inferiority randomized controlled trial.
Testelmans, D; Papadopoulos, D; Kalkanis, A; Jacobs, A; Van Hende, F; Vandebotermet, M; Belge, C; Buyse, B.
Afiliação
  • Testelmans D; Department of Pneumology, University Hospitals Leuven, Leuven, Belgium.
  • Papadopoulos D; Laboratory of Respiratory Diseases and Thoracic Surgery (BREATHE), Department of Chronic Diseases and Metabolism, KU Leuven, Leuven, Belgium.
  • Kalkanis A; Department of Pneumology, University Hospitals Leuven, Leuven, Belgium.
  • Jacobs A; Department of Pneumology, University Hospitals Leuven, Leuven, Belgium.
  • Van Hende F; Laboratory of Respiratory Diseases and Thoracic Surgery (BREATHE), Department of Chronic Diseases and Metabolism, KU Leuven, Leuven, Belgium.
  • Vandebotermet M; Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.
  • Belge C; Department of Pneumology, AZ St-Dimpna, Geel, Belgium.
  • Buyse B; Department of Pneumology, AZ Groeninge, Kortrijk, Belgium.
J Sleep Res ; 32(5): e13887, 2023 10.
Article em En | MEDLINE | ID: mdl-36944532
ABSTRACT
The present study aimed to evaluate whether titration of fixed continuous positive airway pressure at home using telemonitoring produces patient outcomes equal to auto-adjusting positive airway pressure titration at home for patients with moderate obstructive sleep apnea. Patients were randomized with a 11 allocation ratio to receive either auto-adjusting positive airway pressure titration based on the median of the 95th percentile pressure across seven nights or fixed continuous positive airway pressure titration based on a fixed calculated pressure and specific adaptations after telemonitoring of device data after 3 and 7 nights. The results of the ambulatory titration were evaluated with in-laboratory polysomnography after 2 weeks. We hypothesized that fixed continuous positive airway pressure titration would be non-inferior to auto-adjusting positive airway pressure titration in respect to continuous positive airway pressure adherence at a 3-month follow-up. A non-inferiority margin of -0.75 hr was prespecified. One-hundred and four patients were randomly allocated to fixed continuous positive airway pressure (n = 52) and auto-adjusting positive airway pressure (n = 52) titration. The mean difference and the 95% confidence intervals in continuous positive airway pressure adherence after 3 months between the two arms were 0.80 (-0.08, 1.69) hr. The non-inferiority hypothesis was confirmed as the lower one-sided 97.5% confidence interval for the mean difference was above the prespecified margin. Patients in the fixed continuous positive airway pressure titration arm were titrated at significantly lower pressure level and had a significantly lower amount of average leaks compared with auto-adjusting positive airway pressure-titrated patients, while there was no difference in residual obstructive apnea-hypopnea index on polysomnography. Telemonitoring enables ambulatory continuous positive airway pressure titration with fixed pressure that is non-inferior to ambulatory titration with auto-adjusting pressure in patients with moderate obstructive sleep apnea.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Apneia Obstrutiva do Sono / Pressão Positiva Contínua nas Vias Aéreas Tipo de estudo: Clinical_trials Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Apneia Obstrutiva do Sono / Pressão Positiva Contínua nas Vias Aéreas Tipo de estudo: Clinical_trials Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article