Your browser doesn't support javascript.
loading
A Temporal Instability Measure for fMRI Quality Assurance.
Schmidt, Tim; Nagy, Zoltán.
Afiliação
  • Schmidt T; Laboratory for Social and Neural Systems Research, University of Zurich, Switzerland.
  • Nagy Z; Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Switzerland.
J Magn Reson Imaging ; 59(1): 325-336, 2024 01.
Article em En | MEDLINE | ID: mdl-37141174
ABSTRACT

BACKGROUND:

There exist several fMRI quality assurance measures to assess scanner stability. Because they have practical and/or theoretical limitations, a different and more practical measure for instability would be desirable.

PURPOSE:

To develop and test a sensitive, reliable and widely applicable temporal instability measure (TIM) for fMRI quality assurance. STUDY TYPE Technical development. PHANTOM Spherical gel phantom. POPULATION A total of 120 datasets from a local Philips scanner with two different receive-only head coils (32ch and 8ch, 60 datasets per coil) were collected as well as 29 additional datasets with three different receive-only head coils (20ch, 32ch, and 64ch) from two additional sites with GE (seven runs with 32ch) and Siemens scanners (seven runs with 32ch and Multiband imaging, five runs with 20ch, 32ch, and 64ch) were borrowed. FIELD STRENGTH/SEQUENCE 2D Echo-planar-imaging (EPI). ASSESSMENT A new TIM was proposed that is based on the eigenratio of the correlation coefficient matrix, where each entry of the matrix is a correlation coefficient between two time-points of the time-series. STATISTICAL TESTS Nonparametric bootstrap resampling was used twice to estimate confidence intervals (CI) of the TIM values and to assess the improved sensitivity of this measure. Differences in coil performance were assessed via a nonparametric bootstrap two-sample t-test. P-values <0.05 were considered significant.

RESULTS:

The TIM values ranged between 60 parts-per-million and 10,780 parts-per-million across all 149 experiments. The mean CI was 2.96% and 2.16% for the 120 and 29 fMRI datasets, respectively (the repeated bootstrap analysis gave 2.9% and 2.19%, respectively). The 32ch coils of the local Philips data provided more stable measurements than the 8ch coil (observed two-sample t-values = 26.36, -0.2 and -6.2 for TIM, tSNR, and RDC, respectively. PtSNR = 0.58). DATA

CONCLUSION:

The proposed TIM is particularly useful for multichannel coils with spatially nonuniform receive sensitivity and overcomes several limitations of other measures. As such, it provides a reliable test for ascertaining scanner stability for fMRI experiments. EVIDENCE LEVEL 5. TECHNICAL EFFICACY Stage 1.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Imagem Ecoplanar Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Imagem Ecoplanar Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article