SEC-HPLC analysis of column load and flow-through provides critical understanding of low Protein A step yield.
Protein Expr Purif
; 216: 106418, 2024 Apr.
Article
em En
| MEDLINE
| ID: mdl-38141898
ABSTRACT
For a certain number of mAbs, bispecific antibodies (bsAbs) and Fc-fusion proteins that we worked on, the Protein A capture step experienced low yield (i.e., â¼80%). A previous case study suggested that non-binding aggregate formed in cell culture was the root cause of low Protein A step yield. In the current work, we selected five projects with the low Protein A yield issue to further illustrate this phenomenon. In all cases, existence of non-binding aggregates was confirmed by size-exclusion chromatography-high performance liquid chromatography (SEC-HPLC) analysis of Protein A load and flow-through. In addition, we demonstrated that aggregates failed to bind to Protein A resin mainly due to their large sizes, which prevented them from entering the resin beads. As the data suggested, SEC-HPLC analysis of Protein A load and flow-through, although not a standard procedure, can provide information that is critical for understanding the unexpected performance of Protein A chromatography in cases like those being presented here. Thus, SEC-HPLC analysis of Protein A load and flow-through is highly recommended for antibodies/Fc-fusions suffering from low Protein A yield.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Anticorpos Biespecíficos
/
Técnicas de Cultura de Células
Idioma:
En
Ano de publicação:
2024
Tipo de documento:
Article