Comparison of photorefraction by Plusoptix A12 and cycloplegic autorefraction in children.
Ghadimi, Hadi; Nikdel, Mojgan; Suh, Donny W.
Afiliação
  • Ghadimi H; Private Ophthalmology Practice, Atieh Medical Center, Rasht, 4144694198, Guilan, Iran.
  • Nikdel M; Private Ophthalmology Practice, Atieh Medical Center, Rasht, 4144694198, Guilan, Iran. nikdel.mojgan@gmail.com.
  • Suh DW; Department of Ophthalmology and Visual Science, Gavin Herbert Eye Institute, University of California at Irvine, Irvine, CA, USA.
BMC Ophthalmol ; 24(1): 179, 2024 Apr 19.
Article em En | MEDLINE | ID: mdl-38641836
ABSTRACT

BACKGROUND:

Plusoptix photoscreeners are capable of measuring refractive errors of children from 1 meter distance, without cyloplegia. We aimed to compare refractive data obtained from the newest version of Plusoptix (model 12) with cycloplegic autorefraction.

METHODS:

We examined 111 consecutive children aged 3-7 years first by Plusoptix A12C under manifest condition and subsequently for cycloplegic refraction by Topcon KR-1 tabletop autorefractometer. Sphere, spherical equivalent, cylinder and axis of astigmatism measured by the two methods were analyzed to determine correlation, agreement and differences.

RESULTS:

Binocular examination of 111 children aged 4.86±1.27 years revealed good agreement between refractive data obtained by Plusoptix and cycloautorefraction, according to Bland-Altman plots. Significant (p < 0.001) and strong correlation was found between all refractive measurements (Pearson's r value of 0.707 for sphere, 0.756 for pherical equivalent, and 0.863 for cylinder). Plusoptix mean sphere, spherical equivalent and cylinder were 1.22, 0.56, and -1.32 D, respectively. Corresponding values for cycloautorefraction were 1.63, 1.00, and -1.26 D. The difference between axis of cylinder measured by the two methods was < 10° in 144 eyes (64.9%).

CONCLUSIONS:

Considering the significant agreement and correlation between Plusoptix photoscreener and cycloplegic autorefraction, the need for cycloplegic drops in refractive examination of children may be obviated. The mean difference between cylinder measurements are considerably trivial (0.06 D), but sphere is approximately 0.4 D underestimated by Plusoptix compared to cycloautorefraction, on average.