Performance and Safety of the Extravascular Implantable Cardioverter-Defibrillator Through Long-Term Follow-Up: Final Results From the Pivotal Study.
Friedman, Paul; Murgatroyd, Francis; Boersma, Lucas V A; Manlucu, Jaimie; Knight, Bradley P; Clémenty, Nicolas; Leclercq, Christophe; Amin, Anish; Merkely, Béla; Birgersdotter-Green, Ulrika; Chan, Joseph Yat Sun; Biffi, Mauro; Knops, Reinoud Elwin; Engel, Gregory; Carvajal, Ignacio Muñoz; Epstein, Laurence M; Sagi, Venkata; Johansen, Jens Brock; Sterlinski, Maciej; Steinwender, Clemens; Hounshell, Troy; Abben, Richard; Thompson, Amy E; Zhang, Yan; Wiggenhorn, Christopher; Willey, Sarah; Crozier, Ian.
Afiliação
  • Friedman P; Mayo Clinic, Rochester, MN.
  • Murgatroyd F; King's College Hospital, London, England.
  • Boersma LVA; Cardiology Department of St. Antonius Hospital Nieuwegein Netherland; Amsterdam University Medical Centers Amsterdam, Netherlands.
  • Manlucu J; London Health Sciences Centre, Ontario, Canada.
  • Knight BP; Northwestern University, Evanston, IL.
  • Clémenty N; Clinique du Millenaire, Montpellier, France.
  • Leclercq C; CHU de Rennes - Hôpital Pontchaillou, Rennes, France.
  • Amin A; Riverside Methodist Hospital, Columbus, OH.
  • Merkely B; Heart and Vascular Center, Semmelweis University, Budapest, Hungary.
  • Birgersdotter-Green U; University of California San Diego, San Diego, CA.
  • Chan JYS; Prince of Wales Hospital, Chinese University of Hong Kong, Hong Kong, China.
  • Biffi M; Policlinico Sant'Orsola - Malpighi, Bologna, Italy.
  • Knops RE; Amsterdam University Medical Centers Amsterdam, Netherlands.
  • Engel G; Palo Alto Medical Foundation, Redwood City, CA.
  • Carvajal IM; Hospital Universitario Reina Sofía, Córdoba, Spain.
  • Epstein LM; Northwell Health, Manhasset, NY.
  • Sagi V; Baptist Health, Jacksonville, FL.
  • Johansen JB; Odense Universitetshospital, Odense, Denmark.
  • Sterlinski M; I Klinika Zaburzen Rytmu Serca/ Narodowy Instytut Kardiologii, Warszawa, Poland.
  • Steinwender C; Kepler University Hospital, Linz, Austria.
  • Hounshell T; Iowa Heart Center, West Des Moines, IA.
  • Abben R; Cardiovascular Institute of the South, Houma, LA.
  • Thompson AE; Medtronic Inc., Mounds View, MN.
  • Zhang Y; Medtronic Inc., Mounds View, MN.
  • Wiggenhorn C; Medtronic Inc., Mounds View, MN.
  • Willey S; Medtronic Inc., Mounds View, MN.
  • Crozier I; Christchurch Hospital, Christchurch, New Zealand.
Circulation ; 2024 Sep 26.
Article em En | MEDLINE | ID: mdl-39327797
ABSTRACT

BACKGROUND:

Substernal lead placement of the extravascular implantable cardioverter-defibrillator (EV ICD) permits both defibrillation at thresholds similar to those seen with transvenous ICDs and effective antitachycardia pacing (ATP), while avoiding the vasculature and associated complications. The global Pivotal study has shown the EV ICD system to be safe and effective through 6 months, but long-term experience has yet to be published. We aim to report the performance and safety of the EV ICD system throughout the study.

METHODS:

The EV ICD Pivotal study was a prospective, global, single-arm, pre-market clinical study. Individuals with a class I or IIa indication for a single-chamber ICD per guidelines were enrolled. Freedom from major system- or procedure-related complications, as well as appropriate and inappropriate therapy rates, were assessed through 3 years using the Kaplan-Meier method. Anti-tachycardia pacing success was calculated using simple proportions.

RESULTS:

An implant was attempted in 316 patients [25.3% female, 53.8±13.1 years old, 81.6% primary prevention, LVEF 38.9%±15.4%]. Of 299 patients with a successful implant, 24 experienced 82 spontaneous arrhythmic episodes that were appropriately treated with either ATP only (38, 46.3%), shock only (34, 41.5%), or both (10, 12.2%) for a Kaplan-Meier-estimated rate of first any appropriate therapy of 9.2% at 3 years. Antitachycardia pacing was successful in 77.1% (37/48) of episodes, and ATP usage significantly increased from discharge to last follow-up visit (P<0.0001). Shock therapy was successful in 100% (27/27) of discrete, spontaneous ventricular arrhythmias. The inappropriate shock rates at 1 and 3 years were 9.8% and 17.5%, respectively, with P-wave oversensing the predominant cause. No major intraprocedural complications were reported and the estimated freedom from system- or procedure-related major complications was 91.9% at 1 year and 89.0% at 3 years. The most common major complications were lead dislodgement (10 events; n=9 patients, 2.8%), postoperative wound or device pocket infection (n=8, 2.5%), and device inappropriate shock delivery (n=4, 1.3%). Twenty-four system revisions were performed as a result of major complications related to the EV ICD system or procedure.

CONCLUSIONS:

From implant to study completion, the EV ICD Pivotal study demonstrated that a single integrated system with an extravascular lead placed in the substernal space maintains high ATP success, effective defibrillation, and a consistent safety profile.