Your browser doesn't support javascript.
loading
Infinite dilution conductimetry of plasma and urine: correlation with osmolality.
Clin Chim Acta ; 88(1): 177-82, 1978 Aug 15.
Article em En | MEDLINE | ID: mdl-679486
ABSTRACT
The infinite dilution conductivity (IDC) of plasma and urine allows a measurement of the electrolyte content in small samples (5 to 15 microliter). The method was compared to the corrected osmolality (II'p) measured by the freezing-point depression. A linear correlation existed between II'p and the IDC for plasma II'p = 13.10 sigma o,p + 37.00 (n = 46 and r = 0.9949) for urine II'u = 12.75 sigma o,u + 16.56 (n = 85 and r = 0.9504). The measurement of the IDC does not depend on protein concentration and can be used instead of the osmometer methods to determine the total plasma and urine electrolyte content.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Urina / Fenômenos Fisiológicos Sanguíneos / Condutometria Limite: Humans Idioma: En Ano de publicação: 1978 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Urina / Fenômenos Fisiológicos Sanguíneos / Condutometria Limite: Humans Idioma: En Ano de publicação: 1978 Tipo de documento: Article