Your browser doesn't support javascript.
loading
Montrer: 20 | 50 | 100
Résultats 1 - 1 de 1
Filtrer
Plus de filtres








Gamme d'année
1.
Article de Chinois | WPRIM | ID: wpr-619450

RÉSUMÉ

Objective To investigate the necessity of dilution in samples with high concentrations of D-Dimer and optimum dilution multiple.Methods Quality control products and calibration were detected by using Sysmex CS5100 for precision evaluation,including within-batch and between-run precision.Calibration were detected for validation of linear range and clinical reportable.Samples with D-Dimer5 mg/L and FDP>20 μg/mL were also serially diluted and detected to calculated recovery rate.Results Within-batch and between-run coefficients of variation were both less than 3%.Within the scope of 0.207-5.170 mg/L,the linear distribution was fine.The clinical reportable range was 0.207-165.440 mg/L.For samples with D-Dimer5 mg/L and FDP>20 μg/mL,there was obvious antigen excess phenomenon,and gradient dilution was required.Conclusion For samples with D-Dimer>5 mg/L and FDP>20 μg/mL,dilution should be performed to ensure the accuracy of detected results.

SÉLECTION CITATIONS
DÉTAIL DE RECHERCHE