RESUMO
Echocardiogram is vital for the diagnosis of cardiac disease. The heart has complex hemodynamics requiring an advanced ultrasound imaging mode. Cardiac ultrasound vector flow imaging is capable of measuring the actual magnitude and direction of the blood flow velocity, obtaining the quantitative parameters of hemodynamics, and then providing more information for clinical research and diagnosis. This study mainly reviewed several different vector flow imaging techniques for cardiac flow and presented the implementation difficulties, and proposed a diverging wave based high frame rate cardiac ultrasound vector flow imaging. The study discussed the limitation of current ultrasound technology used in the cardiac flow measurement, analyzed and demonstrated the specific reasons for these implementation difficulties and the potential future development.
Assuntos
Velocidade do Fluxo Sanguíneo , Coração/fisiologia , Hemodinâmica , UltrassonografiaRESUMO
The paper described the hemodynamics of blood flow based on fluid mechanics and its corresponding formulas, and revealed the limitation of blood flow velocity measurement for non-laminar flow when using the conventional pulse wave Doppler. The paper demonstrated the calculation of turbulence for blood flow based on velocity directions and quantified the turbulence according to the presented formulas. Two methods were introduced and the simulated results were analyzed. An example using real data based on ultrasound vector flow imaging for calculating the turbulence of blood flow was presented in the end.