We present two cases of misplaced central venous catheters having in common theabsence of free blood return from one lumen immediately after placement. The former is acase of right hydrothorax associated with central venous catheterization with the catheter tipin intra-pleural location. In this case the distal port was never patent. In the latter case therewas an increased aspiration pressure through the middle port due to a catheter looping.The absence of free flow on aspiration from one lumen of a central catheter should not beundervalued. In these circumstances the catheter should not be used and needs to be removed.


Apresentamos dois casos de mau posicionamento de cateter venoso central. Têmem comum a ausência do retorno sanguíneo livre em um dos lúmens imediatamente após acolocac¸ão. O primeiro é um caso de hidrotórax direito associado ao cateterismo venoso central,com a ponta do cateter em localizac¸ão intrapleural. Nesse caso, a porta distal nunca estevepatente. No segundo caso houve um aumento da pressão de aspirac¸ão através da porta medialpor causa da formac¸ão de alc¸a no cateter.A ausência de fluxo livre na aspirac¸ão de um lúmen do cateter central não deve ser subesti-mada. Nessas circunstâncias, o cateter não deve ser usado e deve ser removido.


Subject(s)
Humans , Female , Adult , Aged , Catheterization, Central Venous/adverse effects , Central Venous Catheters/adverse effects , Hydrothorax/etiology , Equipment Failure , Hydrothorax/diagnostic imaging