Maintenance of Shinva PS-100X Medical Low-Temperature Hydrogen Peroxide Plasma Sterilizer:8 case reports / 医疗卫生装备
Chinese Medical Equipment Journal
; (6): 117-120, 2023.
Article
ي Zh
| WPRIM
| ID: wpr-1022879
المكتبة المسؤولة:
WPRO
ABSTRACT
The structure and working principle of Shinva PS-100X Medical Low-Temperature Hydrogen Peroxide Plasma Sterilizer were introduced.Eight cases of the sterilizer failures had their phenomena,causes and elimination described in detail.The importance of daily maintenance and periodical examination was pointed out for the sterilizer.References were provided for clinical engineers to treat similar failures.[Chinese Medical Equipment Journal,2023,44(9):117-120]