Your browser doesn't support javascript.
loading
Research on Constant-current Characteristics of Howland Current Source Used in Bioelectrical Impedance Detection / 生物医学工程学杂志
Article ي Zh | WPRIM | ID: wpr-266684
المكتبة المسؤولة: WPRO
ABSTRACT
This study aims to analyze and improve Howland current source circuit and to study the constant current source of alternating current with high output impedance and high stability. A simulation study was carried out on the constant-current characteristics of Howland current source from two aspects of resistance match value and the selection of Op amp parameters, and then the output impedance was analyzed. The simulation experiment showed that when it was with the best matched resistance, the constant-current characteristics of Howland current source was better. Op amp parameters could affect the constant-current characteristics of Howland current source. In Howland current source circuit, the current source after parameter optimization had better output impedance and load capacity. The results showed that there was the best matched resistance in the selection by Howland current source. The Op amp with large broadband, high slew rate and open loop gain, and wide range of power supply voltage is more suitable for the bioelectrical impedance detection circuit.
الموضوعات
النص الكامل: 1 الفهرس: WPRIM الموضوع الرئيسي: Electric Power Supplies / Electric Impedance / Electricity نوع الدراسة: Diagnostic_studies اللغة: Zh مجلة: Journal of Biomedical Engineering السنة: 2015 نوع: Article
النص الكامل: 1 الفهرس: WPRIM الموضوع الرئيسي: Electric Power Supplies / Electric Impedance / Electricity نوع الدراسة: Diagnostic_studies اللغة: Zh مجلة: Journal of Biomedical Engineering السنة: 2015 نوع: Article