Accuracy of three-dimensional facial image based on phase measuring profilometry / 中华口腔医学杂志
Chinese Journal of Stomatology
; (12): 37-40, 2007.
Article
ي Zh
| WPRIM
| ID: wpr-292991
المكتبة المسؤولة:
WPRO
ABSTRACT
<p><b>OBJECTIVE</b>Three-dimensional (3-D) facial images are very useful for orthodontic diagnosis. Three-dimensional facial imaging system (DSC-2) based on phase measuring profilometry (PMP) was used to obtain images and tested for accuracy.</p><p><b>METHODS</b>Geometric strips (20 mm in width) was used to test the accuracy of the obtained images. Three-dimensional images of plaster head model with marker of 5 mm x 5 mm were taken by DSC-2 three-dimensional facial imaging system. The difference of measurements was compared between both methods. Measurements were repeated twice by three different doctors.</p><p><b>RESULTS</b>The average value of width in 3-D mode was 19.5 mm. No significant difference was found, compared with actual width (20 mm). There was no significant difference between computer measurement and manual measurement of the plaster head model (P > 0.05).</p><p><b>CONCLUSIONS</b>DSC-2 three-dimensional facial imaging system was accurate and reliable to obtain 3-D facial images.</p>
النص الكامل:
1
الفهرس:
WPRIM
الموضوع الرئيسي:
Photogrammetry
/
Imaging, Three-Dimensional
/
Face
/
Methods
نوع الدراسة:
Guideline
المحددات:
Humans
اللغة:
Zh
مجلة:
Chinese Journal of Stomatology
السنة:
2007
نوع:
Article