ABSTRACT
In this study, we have fabricated a fiber-optic dosimeter for a proton beam therapy dosimetry. We have measured scintillating lights with the various kinds of organic scintillators and selected the BCF-12 as a sensor-tip material due to its highest light output and peak/plateau ratio. To determine the optimum diameter of BCF-12, we have measured scintillating lights according to the energy losses of proton beams in a water phantom. Also, we determined the adequate length of organic scintillator by measuring scintillating lights according to the incident angles of proton beam. Using an optimized fiber-optic dosimeter, we have measured scintillating lights according to the dose rates and monitor units of proton accelerator.
الموضوعات
Key words
النص الكامل: 1 الفهرس: WPRIM الموضوع الرئيسي: Organothiophosphorus Compounds / Protons / Water / Optical Fibers / Proton Therapy / Light اللغة: Ko مجلة: Korean Journal of Medical Physics السنة: 2010 نوع: Article