Your browser doesn't support javascript.
loading
In-vitro electrochemical stability evaluation of a flexible MEMS microelectrode / 中国医疗器械杂志
Article ي Zh | WPRIM | ID: wpr-342898
المكتبة المسؤولة: WPRO
ABSTRACT
Three-electrode testing method was used to investigate the effect of temperature on electrode impedance, and the pH shifts in saline solution resulting from the electrical stimulation were also determined. Experiments in PBS (phosphate buffered solution) solution showed that the electrode impedance was almost invariable at the human body temperature range (35 degrees C-40 degrees C). And the experiments in unbuffered saline solution showed that pH shifts decreased from 0.03 to 0.005 when the frequency of biphasic charged-balanced pulses increased from 1 Hz to 100 Hz. Even stimulated by monophasic pulses (frequency is 15 Hz, amplitude is 50 microA), the stimulus-induced pH shift of electrode only varies 0.15 (anodic pulse current increased 0.15 and cathodic pulse current decreased 0.15).
الموضوعات
النص الكامل: 1 الفهرس: WPRIM الموضوع الرئيسي: Prostheses and Implants / Materials Testing / Electrochemistry / Eye, Artificial / Microelectrodes اللغة: Zh مجلة: Zhongguo Yi Liao Qi Xie Za Zhi السنة: 2013 نوع: Article
النص الكامل: 1 الفهرس: WPRIM الموضوع الرئيسي: Prostheses and Implants / Materials Testing / Electrochemistry / Eye, Artificial / Microelectrodes اللغة: Zh مجلة: Zhongguo Yi Liao Qi Xie Za Zhi السنة: 2013 نوع: Article