Comparison of the distribation of doses calculated with Monte Carlo N-particle transport code and those practically measured by 60Co therapy facility / 中华放射医学与防护杂志
Article ي Zh | WPRIM | ID: wpr-412731
المكتبة المسؤولة: WPRO
ABSTRACT
Objective To discuss the feasibility of Monte Carlo N-particle transport code(MCNP)simulated calculation.Methods The calculation in water phantom was contrasted with the practical measurements and the reported values using the percent depth dose(PDD)curve and normal peak scatter factor.Results There Was no significant difference between calculated and measured results in the 10 cm×10 cm field(t=-0.41,P>0.05),however,there were significant differences in the 5 cm×5 cm field(t=7.2,P<0.05)and in the 12 cm×12 cm field(t=-4.6,P<0.05).There was no significant difierence between the calculated results and the reported values(t=-1.91,P>0.05).In the same radiation field,the PDD decreased as the depth increased,but increased as the size of the radiation field increased at the same depth.PDD and normal peak scatter factor were both important parameters for calculation of prescribed dose.Conclusions It is possible to establish a set of accurate and comprehensive percent depth doses and normal peak scatter factor parameters so as to provide the basis of clinical use, quality assurance and quality control for radiotherapy.