Dosimetric analysis of 3D-printing non-coplanar template combined with CT-guided125I seed implantation for the treatment of spinal metastasis / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology
; (6): 122-125, 2020.
Article
ي Zh
| WPRIM
| ID: wpr-799443
المكتبة المسؤولة:
WPRO
ABSTRACT
Objective@#To compare the preoperative and postoperative dosimetric parameters in the treatment of spinal metastasis, and to verify the accuracy of 3D-printing non-coplanar template (3D-PNCT) combined with CT-guided 125I seed implantation for the treatment of spinal metastasis.@*Methods@#The treatment plans of 7 patients with spinal metastasis (9 lesions) from 2016 to 2018 receiving 3D-PNCT in combination with CT-guided 125I seed implantation were retrospectively analyzed. The dosimetric parameters including homogeneity index (HI), conformal index (CI), external index (EI), dose of 90% target volume(D90), mPD, volume percent of 100%, 150%, and 200% prescribed dose V100、V150、V200 and D2cm3 of spinal cord were compared before and after operation. The british columbia cancer ageny particle implantation quality evaluation standard was applied to evaluate the quality of implantation.@*Results@#The HI, EI and CI, D90, mPD, V100, V150, V200 and D2cm3 of spinal cord did not significantly differ before and after the plan (all P>0.05). Five were evaluated as excellent and 4 were assessed as good.@*Conclusion@#The postoperative dosimetric parameters of 3D-PNCT combined with CT guided 125I seed implantation of spinal metastasis are basically consistent with preoperative dosimetric parameters. The postoperative plans are evaluated as excellent or good, suggesting that the technology has a good therapeutic accuracy in the treatment of spinal metastasis.
النص الكامل:
1
الفهرس:
WPRIM
اللغة:
Zh
مجلة:
Chinese Journal of Radiation Oncology
السنة:
2020
نوع:
Article