Cochlear electrode array misplacement into the superior semicircular canal: a case report and literature review / 临床耳鼻咽喉头颈外科杂志
Journal of Clinical Otorhinolaryngology Head and Neck Surgery
; (12): 310-312, 2023.
Article
ي Zh
| WPRIM
| ID: wpr-982739
المكتبة المسؤولة:
WPRO
ABSTRACT
Electrode array misplacement is a rare complication of cochlear implant. This article reports an 11-year-old boy who was mistakenly implanted the cochlear electrode array into the superior semicircular canal during the initial cochlear implant. After the diagnosis was confirmed, he underwent a second cochlear implant and the electrode array were successfully implanted into the cochlea. This article conducted a systematic review of the literature on electrode array misplacement, and the causes of electrode array misplacement were analyzed from different implantation position.
Key words
النص الكامل:
1
الفهرس:
WPRIM
الموضوع الرئيسي:
Reoperation
/
Semicircular Canals
/
Cochlear Implants
/
Cochlea
/
Cochlear Implantation
/
Electrodes, Implanted
المحددات:
Child
/
Humans
/
Male
اللغة:
Zh
مجلة:
Journal of Clinical Otorhinolaryngology Head and Neck Surgery
السنة:
2023
نوع: